Real Time Nucleic Acid Amplification System (OOI) — FDA decisions, comparisons and evidence

The 2021–2025 analysis contains 3 selected substantially-equivalent FDA decisions under primary product code OOI. Use it to compare documented submissions and evidence with your product. A separate 2026 update contains 0 decisions through 2026-09-27. The latest selected decision across the acquired endpoint is dated 2025-03-20. Neither set establishes a suitable predicate or the route for your device. The complete-year window has 3 decisions; the separate all-observed-date count is 14. These denominators describe different date ranges.

Evidence retrieved 2026-10-07. Source versions and topic-specific limits are listed below.

Recorded scope and its declared regulatory reference

An exact FDA-record regulation_number match to one acquired Title 21 section. Records with the same normalized scope definition, class, regulation and recorded submission/GMP flags are grouped. Sparse definitions and generic exemption boilerplate do not qualify. No text-similarity equivalence or legal applicability is inferred.

What scope does the FDA record describe?

The system is a clinical multiplex instrument intended to measure and sort multiple signals generated my multiple probes, intercalating dyes, or other ligands in an assay from a clinical sample. Signals may be generated by fluorescence or other phenomena and may be measured using filters on a photodiode or other detector. It may integrate sample and/or reagent handling, amplification, dedicated instrument control, data acquisition software, raw data storage mechanisms and other essential hardware components along with the signal reader unit. The system is used with specific assays to comprise an assay test system.

Compare the proposed indication, user, anatomy, technology and operating principle with this actual scope; record differences and unresolved facts.

Cited source

Which current section does the record cite?

§ 862.2570 Instrumentation for clinical multiplex test systems. (a) Identification. Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals. (b) Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document. [70 FR 11868, Mar. 10, 2005, as amended at 84 FR 71797, Dec. 30, 2019]

Read the identification, classification, conditions and referenced limitations in the cited section. A numeric reference match is not a buyer classification or exemption determination.

Cited source

Historical FDA comparison

2021–2025: five complete calendar years

Primary code OOI

All statistics in this section use decisions dated 2021-01-01 to 2025-12-31. The partial 2026 update below is excluded from these distributions.

3Selected decisions in these five years
WithheldMedian receipt-to-decision calendar days
3Valid date pairs in the distribution

There are 3 valid recent date pairs. Distribution summaries require at least 20; older records are not substituted for a current benchmark.

20210
20221
20231
20240
20251

Compare like submission types

Recorded typeDecisionsValid date pairsMedian calendar daysMiddle 50%
Special11Withheld: n < 20—
Traditional22Withheld: n < 20—

Different submission types and evidence packages are not interchangeable. These selected records do not establish that a particular route is available for your product.

Statistical source: FDA openFDA 510(k) decision dataset. Partition 1 Download the identified records and dates (CSV).

Filters, exclusions and reproducible calculation

Deduplicate by official submission ID across the complete current manifest partitions. Select exact primary product code, K-number format and the stated SE decision codes with valid dates. Recent distributions use the five complete calendar years preceding the latest valid endpoint decision year. Partial-year counts compare equal January-to-cutoff periods. Quartiles use linear interpolation at (n-1)*p and are withheld below 20 valid date pairs. The analysis n describes only its declared complete-year window; selected_se_n separately counts all selected recorded dates through the cutoff.

14 selected records across all observed dates; 0 other/invalid identifier or decision-date records excluded. 0 missing or invalid date pairs in the five-year window.

Date fields: date_received → decision_date. Quantiles: Hyndman–Fan type 7: linear interpolation at (n − 1) × p; displayed to one decimal; n ≥ 20 valid pairs.

Receipt-to-decision calendar elapsed time includes time outside active FDA review; it is not FDA review time, a promised project timeline or an estimate of future clearance. This selected recorded cohort does not include all applications or establish a success probability, predicate suitability, market size, current market availability or legal authorization for another product.

Calculation fda-buyer-research-3 · database cutoff 2026-09-27. CSV rows identify each official K-number, cohort, date pair, exclusion reason and source version.

Separate partial-year update

2026 decisions through 2026-09-27

0 selected decisions from 2026-01-01 to 2026-09-27. The equivalent previous-year period contains 1 decisions through 2025-09-27. These counts describe the records; they are not market growth or submission success rates.

Named records to investigate

Latest decisions across the database

These dated records may come from 2026 or earlier years. They are a separate investigation list, not the five-year statistical cohort. Compare the actual indications and technology before considering a record as a comparator.

Official record / deviceRecorded applicantDecisionRecorded typeCalendar days
K243922 ↗RevogeneMeridian Bioscience, Inc.2025-03-20Traditional90
K222779 ↗RevogeneMeridian Bioscience, Inc.2023-01-26Special134
K220480 ↗RevogeneMeridian Bioscience, Inc.2022-07-11Traditional143
K170558 ↗revogeneGenepoc, Inc.2017-05-25Traditional90
K161495 ↗ARIES MI System, ARIES HSV 1&2 Assay Cassettes-Carton of 24 (IVD), ARIES HSV 1&2 Assay Protocol File Kit (IVD)Luminex Corporation2016-06-30Special29
K160517 ↗ARIES SystemLuminex Corporation2016-04-12Special48
K151917 ↗ARIES SystemLuminex Corporation2015-10-06Traditional85
K151690 ↗Alere i Instrument, Alere i Influenza A & B, Alere i Strep AAlere Scarborough, Inc.2015-07-16Special23

Supporting documents actually acquired

Go beyond the database row

Read the source context, then compare the evidence with your design. Topic locations below are text matches, including possible limitations or negative statements; they are not a mandatory test list.

The documents are a bounded sample of acquired summaries, not complete evidence coverage of the cohort.

Recorded classification context

Known context: US. Match the intended use and design with the recorded category before treating it as applicable.

Source factRecorded value
FDA product codeOOI [1]
Generic device categoryReal Time Nucleic Acid Amplification System [1]
Recorded scopeThe system is a clinical multiplex instrument intended to measure and sort multiple signals generated my multiple probes, intercalating dyes, or other ligands in an assay from a clinical sample. Signals may be generated by fluorescence or other phenomena and may be measured using filters on a photodiode or other detector. It may integrate sample and/or reagent handling, amplification, dedicated instrument control, data acquisition software, raw data storage mechanisms and other essential hardware components along with the signal reader unit. The system is used with specific assays to comprise an assay test system. [1]
Recorded class2 [1]
Regulation862.2570 [1]
Medical specialtyClinical Chemistry [1]

Build a comparison that explains the differences

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Which specific part of the recorded scope fits or differs?

    Put your proposed label and design beside the quoted definition. Record matching facts, differences and missing facts separately; naming the category alone cannot resolve scope.

    Useful evidence: Proposed indication/design and a definition-to-product comparison with source locators.

  2. Have the cited section and its limitations been reviewed?

    Record the applicable paragraph, conditions and cross-referenced limitations after specialist review. Keep a claimed exemption separate from actual establishment, listing and quality-system responsibilities.

    Useful evidence: Dated classification/route rationale and the current provisions relied on, with unresolved conditions.

  3. Compare your intended use with a named decision

    Choose a named record above. Put your proposed claim beside its actual indications-for-use statement. Record different patients, users, anatomy, settings and output claims; do not treat a shared code as proof of equivalence.

    Useful evidence: Your draft indications for use + the selected official summary and its exact page.

  4. Explain the technology and evidence differences

    For each comparison, record the different materials, hardware, software functions and operating conditions. Link each difference to existing evidence or an unresolved evaluation task.

    Useful evidence: A three-column matrix: comparator fact / your design fact / evidence or unresolved gap.

  5. Prepare a scope-based schedule without a sparse timing benchmark

    This recent cohort has fewer than 20 valid date pairs, so it supplies no median or percentile benchmark. Ask for a schedule based on actual preparation, evidence gaps, interactions and response assumptions; keep any historical decision context separately dated.

    Useful evidence: Document/test readiness, unresolved route/evidence tasks and the assumptions behind the specialist’s proposed sequence.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — device classification records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 936de5f5d47b9293b702fc32512a79173969fb788c74a63d9b91abe73f3b86a5

  • [1] device-classification-0001-of-0001.json:results[5683] · record e56ff2df4b31f063d2b737f36c8de2c8a4d616fc24848a79e48bf0245e1fe41f

FDA 510(k) summary — K222779 ↗

Fri, 27 Jan 2023 22:39:34 GMT · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 27d3d2b97123ef8cf5b7dce28d4491a6dc1c21ed4587ebbfbddca3ff1243803f

  • [3] PDF page 7 · record 2e5080d056f104ec0f1371d9b333fc86ff87ef69da6e3d9b964ee1ce02a5fd46

FDA 510(k) summary — K220480 ↗

Wed, 13 Jul 2022 00:00:04 GMT · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 7af0f57dfc2374750ee232cec71fb6c54c474f7269a9f71db2156b6cd973e6f3

  • [4] PDF page 7 · record 9dae6c3145b73bdba6914b3493afefb54527e077ad0b819a279d46ff7f93780f