FDA record research

Pad, Menstrual, Scented, Scented-Deodorized (HHL) — FDA decisions, comparisons and evidence

The acquired FDA endpoint has no selected substantially-equivalent K-number decisions under primary product code HHL during 2021–2025. Its latest selected record across the endpoint is dated 1996-10-10. Use that dated record only as historical context. Check the recorded scope and current route separately; these data do not provide a recent timing benchmark. The complete-year window has 0 decisions; the separate all-observed-date count is 57. These denominators describe different date ranges.

Evidence retrieved 2026-10-06. Source versions and topic-specific limits are listed below.

Recorded scope and its declared regulatory reference

An exact FDA-record regulation_number match to one acquired Title 21 section. Records with the same normalized scope definition, class, regulation and recorded submission/GMP flags are grouped. Sparse definitions and generic exemption boilerplate do not qualify. No text-similarity equivalence or legal applicability is inferred.

What scope does the FDA record describe?

Scented or scented-deodorized menstrual pads used to absorb menstrual or other vaginal discharge, including those intended as intralabial pads or reusable menstrual pads. These are made with materials that do not meet the requirements for Class 1 per 21 CFR 884.5425(b).

Compare the proposed indication, user, anatomy, technology and operating principle with this actual scope; record differences and unresolved facts.

Cited source

Which current section does the record cite?

§ 884.5425 Scented or scented deodorized menstrual pad. (a) Identification. A scented or scented deodorized menstrual pad is a device that is a pad made of cellulosic or synthetic material which is used to absorb menstrual or other vaginal discharge. It has scent (i.e., fragrance materials) added for aesthetic purposes (scented menstrual pad) or for deodorizing purposes (scented deodorized menstrual pad). This generic type of device includes sterile scented menstrual pads used for medically indicated conditions, but does not include menstrual pads treated with added antimicrobial agents or other drugs. (b) Classification. (1) Class I (general controls) for menstrual pads made of common cellulosic and synthetic material with an established safety profile. The devices subject to this paragraph (b)(1) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9. This exemption does not include the intralabial pads and reusable menstrual pads. (2) Class II (special controls) for scented or scented deodorized menstrual pads made of materials not described in paragraph (b)(1). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9. [45 FR 12684, Feb. 26, 1980, as amended at 45 FR 51185, Aug. 1, 1980; 61 FR 67714, Dec. 24, 1996; 66 FR 38809, July 25, 2001; 84 FR 71816, Dec. 30, 2019]

Read the identification, classification, conditions and referenced limitations in the cited section. A numeric reference match is not a buyer classification or exemption determination.

Cited source

Historical FDA comparison

2021–2025: five complete calendar years

Primary code HHL

All statistics in this section use decisions dated 2021-01-01 to 2025-12-31. The partial 2026 update below is excluded from these distributions.

0Selected decisions in these five years
WithheldMedian receipt-to-decision calendar days
0Valid date pairs in the distribution

There are 0 valid recent date pairs. Distribution summaries require at least 20; older records are not substituted for a current benchmark.

20210
20220
20230
20240
20250

Statistical source: FDA openFDA 510(k) decision dataset. Partition 1 Download the identified records and dates (CSV).

Filters, exclusions and reproducible calculation

Deduplicate by official submission ID across the complete current manifest partitions. Select exact primary product code, K-number format and the stated SE decision codes with valid dates. Recent distributions use the five complete calendar years preceding the latest valid endpoint decision year. Partial-year counts compare equal January-to-cutoff periods. Quartiles use linear interpolation at (n-1)*p and are withheld below 20 valid date pairs. The analysis n describes only its declared complete-year window; selected_se_n separately counts all selected recorded dates through the cutoff.

57 selected records across all observed dates; 0 other/invalid identifier or decision-date records excluded. 0 missing or invalid date pairs in the five-year window.

Date fields: date_received → decision_date. Quantiles: Hyndman–Fan type 7: linear interpolation at (n − 1) × p; displayed to one decimal; n ≥ 20 valid pairs.

Receipt-to-decision calendar elapsed time includes time outside active FDA review; it is not FDA review time, a promised project timeline or an estimate of future clearance. This selected recorded cohort does not include all applications or establish a success probability, predicate suitability, market size, current market availability or legal authorization for another product.

Calculation fda-buyer-research-3 · database cutoff 2026-09-27. CSV rows identify each official K-number, cohort, date pair, exclusion reason and source version.

Separate partial-year update

2026 decisions through 2026-09-27

0 selected decisions from 2026-01-01 to 2026-09-27. The equivalent previous-year period contains 0 decisions through 2025-09-27. These counts describe the records; they are not market growth or submission success rates.

Named records to investigate

Latest decisions across the database

These dated records may come from 2026 or earlier years. They are a separate investigation list, not the five-year statistical cohort. Compare the actual indications and technology before considering a record as a comparator.

Official record / deviceRecorded applicantDecisionRecorded typeCalendar days
K962962 ↗DEODORANT PANTY LINERParagon Trade Brands, Inc.1996-10-10Traditional72
K962954 ↗DEODORANT MAXI PADParagon Trade Brands, Inc.1996-10-10Traditional72
K962953 ↗DEODORANT PANTY SHIELDParagon Trade Brands, Inc.1996-10-10Traditional72
K961119 ↗CAREFREE PANTYLINERSPersonal Products Co.1996-04-17Traditional27
K913238 ↗ALWAYS, UNSCENTED MENSTRUAL PADProcter & Gamble Co.1991-09-23Traditional63
K913237 ↗ALWAYS, SCENTED/SCENTED DEODORIZED MENSTRUAL PADProcter & Gamble Co.1991-09-23Traditional63
K891473 ↗TAMPAX ULTRAFIT PANTY SHIELDS/DEDORANT-MODIFICATIOTambrands, Inc.1989-06-07Traditional83
K890655 ↗NEW FREEDOM THIN DEODORANT MAXI PADSKimberly-Clark Corp.1989-05-01Traditional82

Recorded classification context

Known context: US. Match the intended use and design with the recorded category before treating it as applicable.

Source factRecorded value
FDA product codeHHL [1]
Generic device categoryPad, Menstrual, Scented, Scented-Deodorized [1]
Recorded scopeScented or scented-deodorized menstrual pads used to absorb menstrual or other vaginal discharge, including those intended as intralabial pads or reusable menstrual pads. These are made with materials that do not meet the requirements for Class 1 per 21 CFR 884.5425(b). [1]
Recorded class2 [1]
Regulation884.5425 [1]
Medical specialtyObstetrics/Gynecology [1]

Build a comparison that explains the differences

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Which specific part of the recorded scope fits or differs?

    Put your proposed label and design beside the quoted definition. Record matching facts, differences and missing facts separately; naming the category alone cannot resolve scope.

    Useful evidence: Proposed indication/design and a definition-to-product comparison with source locators.

  2. Have the cited section and its limitations been reviewed?

    Record the applicable paragraph, conditions and cross-referenced limitations after specialist review. Keep a claimed exemption separate from actual establishment, listing and quality-system responsibilities.

    Useful evidence: Dated classification/route rationale and the current provisions relied on, with unresolved conditions.

  3. Separate historical research from a recent comparator

    No selected decision falls in the complete-year window. Any named records shown are separately dated historical or current-year research. Obtain their actual indications and assess the current route before using a comparator; the absence of a recent record does not establish an exemption or a route.

    Useful evidence: Proposed indication/design, the dated named record if available, and a current route/comparator research rationale.

  4. Explain the technology and evidence differences

    For each comparison, record the different materials, hardware, software functions and operating conditions. Link each difference to existing evidence or an unresolved evaluation task.

    Useful evidence: A three-column matrix: comparator fact / your design fact / evidence or unresolved gap.

  5. Prepare a scope-based schedule without a sparse timing benchmark

    This recent cohort has fewer than 20 valid date pairs, so it supplies no median or percentile benchmark. Ask for a schedule based on actual preparation, evidence gaps, interactions and response assumptions; keep any historical decision context separately dated.

    Useful evidence: Document/test readiness, unresolved route/evidence tasks and the assumptions behind the specialist’s proposed sequence.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — device classification records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 936de5f5d47b9293b702fc32512a79173969fb788c74a63d9b91abe73f3b86a5

  • [1] device-classification-0001-of-0001.json:results[6200] · record 7e1bdf10a63240021f1cf930c8af55119145657213bdd5c9578b994697da198a

Prepare an editable project brief

Confirm the facts, scope and contact preference before sharing your project. Preparing this page sends no provider outreach.

Choose work packages to discuss

Compare FDA 510(k) Submission Services

Find FDA US Agent Services | Compare & Get Quotes · FDA QMSR Transition & Inspection Readiness (ISO 13485 Alignment)