Culture Plate Imaging System For Qualitative Assessment Of Resistant Organisms (QQY) — FDA decisions, comparisons and evidence

The 2021–2025 analysis contains 2 selected substantially-equivalent FDA decisions under primary product code QQY. Use it to compare documented submissions and evidence with your product. A separate 2026 update contains 0 decisions through 2026-09-27. The latest selected decision across the acquired endpoint is dated 2023-05-04. Neither set establishes a suitable predicate or the route for your device. The complete-year window has 2 decisions; the separate all-observed-date count is 2. These denominators describe different date ranges.

Evidence retrieved 2026-10-07. Source versions and topic-specific limits are listed below.

Recorded scope and its declared regulatory reference

An exact FDA-record regulation_number match to one acquired Title 21 section. Records with the same normalized scope definition, class, regulation and recorded submission/GMP flags are grouped. Sparse definitions and generic exemption boilerplate do not qualify. No text-similarity equivalence or legal applicability is inferred.

What scope does the FDA record describe?

Culture plate imaging system for qualitative assessment of resistant organisms is an automated system to assess the presence or absence of colony growth on solid chromogenic culture media.

Compare the proposed indication, user, anatomy, technology and operating principle with this actual scope; record differences and unresolved facts.

Cited source

Which current section does the record cite?

§ 866.2190 Automated image assessment system for microbial colonies on solid culture media. (a) Identification. An automated image assessment system for microbial colonies on solid culture media is a system that is intended to assess the presence or absence of microbial colonies on solid microbiological culture medium, and to interpret their number, and phenotypic and morphologic characteristics through analysis of two dimensional digital images as an aid in diagnosis of infectious disease. (b) Classification. Class II (special controls). The special controls for this device are: (1) Premarket notification submissions must include a detailed description of the device, including the technology employed, components and software modules, as well as a detailed explanation of the result algorithms and any expert rules that are used to assess colony characteristics and enumerate colonies from image capture through end result. (2) Premarket notification submissions must include detailed documentation of the analytical studies performed to characterize device performance to support the intended use, as appropriate. (3) Premarket notification submissions must include detailed documentation from clinical studies performed on a population that is consistent with the intended use population. (i) The clinical studies must establish the device performance based on comparison to results obtained by an acceptable reference method, as appropriate. (ii) The clinical study documentation must include the study protocol with a predefined statistical analysis plan and the final report documenting support for the Indications for Use and the results of the statistical analysis, as appropriate. (4) Premarket notification submissions must include detailed documentation for device software, including but not limited to software applications and hardware based components that incorporate software, and any decision-making thresholds used to generate results for the device. If a part of a Total Laboratory Automation System, the premarket notification submission must include detailed documentation addressing the instrument and software system integration. (5) Premarket notification submissions must include detailed documentation of appropriate instructions for use regarding the intended user's device quality control procedures for the instrument system and components, as appropriate. (6) The 21 CFR 809.10 compliant device labeling must include: (i) Detailed user instructions to mitigate the risk of failure to operate the instrument correctly. (ii) A detailed explanation of the interpretation of results and limitations regarding the need for review of culture plates by a qualified microbiologist, as appropriate. (iii) A summary of performance data obtained from the analytical studies used to support device performance, as appropriate. (iv) A summary of performance data obtained from clinical studies performed on a population that is consistent with the intended use population, as appropriate. (7) Under 21 CFR 820.10(c) design and development, device manufacturers must, as appropriate: (i) Conduct human factors/usability validation testing with the final version of the labeling and related materials to adequately mitigate the risk of failure to operate the instrument correctly. (ii) Document a device training program that will be offered to the end user to adequately mitigate the risk of failure to operate the instrument correctly. [82 FR 47969, Oct. 16, 2017, as amended at 90 FR 55982, Dec. 4, 2025]

Read the identification, classification, conditions and referenced limitations in the cited section. A numeric reference match is not a buyer classification or exemption determination.

Cited source

Historical FDA comparison

2021–2025: five complete calendar years

Primary code QQY

All statistics in this section use decisions dated 2021-01-01 to 2025-12-31. The partial 2026 update below is excluded from these distributions.

2Selected decisions in these five years
WithheldMedian receipt-to-decision calendar days
2Valid date pairs in the distribution

There are 2 valid recent date pairs. Distribution summaries require at least 20; older records are not substituted for a current benchmark.

20211
20220
20231
20240
20250

Statistical source: FDA openFDA 510(k) decision dataset. Partition 1 Download the identified records and dates (CSV).

Filters, exclusions and reproducible calculation

Deduplicate by official submission ID across the complete current manifest partitions. Select exact primary product code, K-number format and the stated SE decision codes with valid dates. Recent distributions use the five complete calendar years preceding the latest valid endpoint decision year. Partial-year counts compare equal January-to-cutoff periods. Quartiles use linear interpolation at (n-1)*p and are withheld below 20 valid date pairs. The analysis n describes only its declared complete-year window; selected_se_n separately counts all selected recorded dates through the cutoff.

2 selected records across all observed dates; 0 other/invalid identifier or decision-date records excluded. 0 missing or invalid date pairs in the five-year window.

Date fields: date_received → decision_date. Quantiles: Hyndman–Fan type 7: linear interpolation at (n − 1) × p; displayed to one decimal; n ≥ 20 valid pairs.

Receipt-to-decision calendar elapsed time includes time outside active FDA review; it is not FDA review time, a promised project timeline or an estimate of future clearance. This selected recorded cohort does not include all applications or establish a success probability, predicate suitability, market size, current market availability or legal authorization for another product.

Calculation fda-buyer-research-3 · database cutoff 2026-09-27. CSV rows identify each official K-number, cohort, date pair, exclusion reason and source version.

Separate partial-year update

2026 decisions through 2026-09-27

0 selected decisions from 2026-01-01 to 2026-09-27. The equivalent previous-year period contains 0 decisions through 2025-09-27. These counts describe the records; they are not market growth or submission success rates.

Named records to investigate

Latest decisions across the database

These dated records may come from 2026 or earlier years. They are a separate investigation list, not the five-year statistical cohort. Compare the actual indications and technology before considering a record as a comparator.

Official record / deviceRecorded applicantDecisionRecorded typeCalendar days
K213280 ↗BD Kiestra Methicillin-resistant Staphylococcus aureus (MRSA) Application, BD Kiestra MRSA AppBd Kiestra B.V.2023-05-04Traditional580
K200839 ↗APAS Independence with IC Chromogenic MRSA BD Analysis Module; APAS Independence with IC Chromogenic MRSA TFS/S Analysis ModuleClever Culture Systems2021-10-28Traditional576

Supporting documents actually acquired

Go beyond the database row

Read the source context, then compare the evidence with your design. Topic locations below are text matches, including possible limitations or negative statements; they are not a mandatory test list.

The documents are a bounded sample of acquired summaries, not complete evidence coverage of the cohort.

Recorded classification context

Known context: US. Match the intended use and design with the recorded category before treating it as applicable.

Source factRecorded value
FDA product codeQQY [1]
Generic device categoryCulture Plate Imaging System For Qualitative Assessment Of Resistant Organisms [1]
Recorded scopeCulture plate imaging system for qualitative assessment of resistant organisms is an automated system to assess the presence or absence of colony growth on solid chromogenic culture media. [1]
Recorded class2 [1]
Regulation866.2190 [1]
Medical specialtyMicrobiology [1]

Build a comparison that explains the differences

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Which specific part of the recorded scope fits or differs?

    Put your proposed label and design beside the quoted definition. Record matching facts, differences and missing facts separately; naming the category alone cannot resolve scope.

    Useful evidence: Proposed indication/design and a definition-to-product comparison with source locators.

  2. Have the cited section and its limitations been reviewed?

    Record the applicable paragraph, conditions and cross-referenced limitations after specialist review. Keep a claimed exemption separate from actual establishment, listing and quality-system responsibilities.

    Useful evidence: Dated classification/route rationale and the current provisions relied on, with unresolved conditions.

  3. Compare your intended use with a named decision

    Choose a named record above. Put your proposed claim beside its actual indications-for-use statement. Record different patients, users, anatomy, settings and output claims; do not treat a shared code as proof of equivalence.

    Useful evidence: Your draft indications for use + the selected official summary and its exact page.

  4. Explain the technology and evidence differences

    For each comparison, record the different materials, hardware, software functions and operating conditions. Link each difference to existing evidence or an unresolved evaluation task.

    Useful evidence: A three-column matrix: comparator fact / your design fact / evidence or unresolved gap.

  5. Prepare a scope-based schedule without a sparse timing benchmark

    This recent cohort has fewer than 20 valid date pairs, so it supplies no median or percentile benchmark. Ask for a schedule based on actual preparation, evidence gaps, interactions and response assumptions; keep any historical decision context separately dated.

    Useful evidence: Document/test readiness, unresolved route/evidence tasks and the assumptions behind the specialist’s proposed sequence.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — device classification records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 936de5f5d47b9293b702fc32512a79173969fb788c74a63d9b91abe73f3b86a5

  • [1] device-classification-0001-of-0001.json:results[5426] · record ea95f9bcc8bea639d17385e73d297506d55db3a2ad59c440717691f22296aa6d

FDA 510(k) summary — K200839 ↗

Fri, 29 Oct 2021 20:00:14 GMT · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot b5265769efa136a04ba354129cd487bef612dcbdefccb27a2d226eb5042c1298

  • [3] PDF page 12 · record 3c5491bdff730edaf85e753f5eaf777f6f2924e03a3544f9d358af142402b78f
  • [4] PDF page 25 · record aefac72b17c351a2300b63de4bcc084cfcc11d80e84bb381b46ea580ece82a2b