Recorded device evidence

Profoject™ Enteral Feeding Syringe (K251585): indications, design and recorded evidence

The acquired K251585 file contains product-specific indications and users, design and operating principle, performance evidence actually described sections. Compare these dated manufacturer statements with your own indication and design before commissioning evidence work. They describe this submission; they do not prescribe your testing, validate performance independently or establish a suitable predicate.

Evidence retrieved 2026-10-07. Source versions and topic-specific limits are listed below.

Computed exact-code research

This decision beside its recorded product-code cohort

The 2021–2025 complete-year cohort contains 10 selected decisions under primary code PNR. 10 have valid reported receipt/decision pairs; 0 pairs are missing or invalid.

No quartiles are displayed: this complete-year cohort has fewer than 20 valid pairs.

The table shows the named decision and up to four latest exact-code research records through 2026-09-27. It can include partial-year records outside the complete-year cohort. The quartiles above use the full stated cohort, not this displayed selection.

Research recordRecorded deviceReceipt → decisionCalendar elapsed daysRecorded submission type
K251585 ↗ · this pageProfoject™ Enteral Feeding Syringe2025-05-23 → 2025-07-1856Traditional
K252483 ↗Enteral Syringe (20ml, 60ml)2025-08-07 → 2026-07-01328Traditional
K253268 ↗Profoject Enteral/Oral Feeding Syringe; Profoject Reusable Enteral/Oral Feeding Syringe (Model A, Model B); Profoject ENFit Adaptor; Profoject Enteral Feeding Syringes Cap2025-09-29 → 2026-04-08191Traditional
K243652 ↗Monoject™ Enteral Syringe with ENFit Connector (401SE); Monoject™ Enteral Syringe with ENFit Connector (403SE); Monoject™ Enteral Syringe with ENFit Connector (406SE); Monoject™ Enteral Syringe with ENFit Connector (412SE); Monoject™ Enteral Syringe with ENFit Connector (435SE); Monoject™ Enteral Syringe with ENFit Connector (460SE)2024-11-26 → 2024-12-1923Special
K231300 ↗Nutricair Enteral Syringe with ENFit2023-05-04 → 2023-10-06155Traditional

Use this comparison: select a record to investigate, read its actual indications and design, and document differences. A shared code or elapsed period cannot establish its relevance to your device.

Calendar elapsed days are not active FDA review time, a promised schedule or a success probability. The selection includes only the stated recorded decisions, not all applications. Latest records can postdate the named historical decision and are contextual research, not its predicates. Shared codes, summaries and observed records do not prove equivalence, current market availability, buyer applicability or predicate suitability.

Calculation, selection and source versions

Reuse the complete-endpoint exact-primary-code, documented SE-decision cohort and its stated cutoff/window. Retain its type-7 quartiles only when at least 20 valid date pairs are present. Independently subtract the named official record’s receipt date from its decision date; missing or negative pairs remain unknown. Show up to four latest exact-code research records, excluding the named ID. Parent analysis hash, source snapshots and calculation versions are retained.

Calculation: fda-named-decision-cohort-context-1; cutoff 2026-09-27; 0 parent records excluded; units record counts; calendar elapsed days.

Source snapshots: 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002.

Read beyond the decision row

Recorded evidence and differences to investigate

The summary is manufacturer-submitted context hosted by FDA. Negative/no-study statements are preserved. Evidence themes are not binding legal requirements.

Short opening excerpts identify the source sections. Read the linked original for the full study context, negative statements, tables and limitations. Public quotations are limited to 120 words per acquired summary; the internal evidence packet retains exact section ranges for review.

Indications and users

K251585 · PDF page 5 ↗ Source 2

The device is intended for use as a dispenser, a measuring device and an
                                                                                               oral fluid transfer device. It is intended to be used to deliver fluids into the
                                                                                               body orally or enterally.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

Design and operating principle

K251585 · PDF page 5 ↗ Source 2

ProfojectTM Enteral Feeding Syringe provided in a variety of sizes from
                                                                                               0.5mL to 60mL.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

Performance evidence actually described

K251585 · PDF page 6 ↗ Source 3

Read this statement in the original PDF; no complete statement fits the short excerpt.

Compare with your product: The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

Extraction selection and limits

Select observed whole-line section headings or explicit colon-delimited inline labels on readable PDF pages, including the explicitly documented heading-spacing variants in this parser version. Preserve the original heading, character ranges and layout; do not repair excerpt text. Exclude letter boilerplate and uncertain pages. Stop at observed table headings; do not infer table columns or scalar results from layout text. Multi-column or suspicious split-numeric excerpts are excluded until reviewed alternate extraction is available. Original layout records remain retained; any alternate text has its own extractor-version locator and requires original-page review. For each dimension, prefer usable versioned alternate text where available, selecting its longest usable excerpt; otherwise select the longest usable original-layout excerpt. Ties preserve source order. Bounded to 64 PDF pages and 3,500 characters per section. This is a selection, not a complete dossier analysis.

Excluded extraction items: {}.

Historical references beside dated FDA recognition metadata

Read each mention in the linked original summary, including statements that testing was not performed. Recognition metadata does not determine your testing needs, chosen edition or transition eligibility.

Reference in this historical summaryDesignation in the acquired FDA exportRecognition / entry dateRecorded transition-expiration dateSources
ISO 20695
Performance Testing
ISO 20695 First edition 2020-03
Enteral feeding systems - Design and testing
9-138 · Complete
06/07/2021
Not recorded in this export rowSummary [3] · Metadata [4]
Read scope and recognition limits ↗
ISO 7886-1
Performance Testing
ISO 7886-1 Second edition 2017-05
Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
6-404 · Complete
06/07/2018
Not recorded in this export rowSummary [3] · Metadata [5]
Read scope and recognition limits ↗
ISO 80369-1
Performance Testing
ISO 80369-1 Second edition 2018-11
Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
5-121 · Complete
01/14/2019
12/17/2028Summary [3] · Metadata [6]
Read scope and recognition limits ↗
ISO 80369-1
Performance Testing
ISO 80369-1 Third edition 2025-10
Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
5-147 · Complete
12/22/2025
Not recorded in this export rowSummary [3] · Metadata [7]
Read scope and recognition limits ↗
ISO 80369-3
Performance Testing
ISO 80369-3 First Edition 2016-07-01
Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications [Including AMENDMENT 1 (2019)].
5-123 · Complete
07/15/2019
Not recorded in this export rowSummary [3] · Metadata [8]
Read scope and recognition limits ↗
ISO 80369-20
Performance Testing
ISO 80369-20 First edition 2015-05-15
Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
5-97 · Complete
08/14/2015
07/02/2028Summary [3] · Metadata [9]
Read scope and recognition limits ↗
ISO 80369-20
Performance Testing
ANSI AAMI ISO 80369-20:2015
Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
5-97 · Complete
08/14/2015
07/02/2028Summary [3] · Metadata [10]
Read scope and recognition limits ↗
ISO 80369-20
Performance Testing
ISO 80369-20 Second edition 2024-11
Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
5-144 · Complete
05/26/2025
Not recorded in this export rowSummary [3] · Metadata [11]
Read scope and recognition limits ↗
ISO 11737-1
Performance Testing
ISO 11737-1 Third edition 2018-01 [Including AMD1:2021]
Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)]
14-577 · Complete
05/30/2022
Not recorded in this export rowSummary [3] · Metadata [12]
Read scope and recognition limits ↗
ISO 10993-7
Performance Testing
ISO 10993-7 Second edition 2008-10-15
Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009) AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)]
2-275 · Complete
07/06/2020
Not recorded in this export rowSummary [3] · Metadata [13]
Read scope and recognition limits ↗
ISO 11135
Performance Testing
ISO 11135 Second edition 2014-07-15
Sterilization of health-care products - Ethylene oxide - Requirements for the development validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]
14-529 · Complete
07/15/2019
Not recorded in this export rowSummary [3] · Metadata [14]
Read scope and recognition limits ↗
ISO 11135
Performance Testing
ANSI AAMI ISO 11135:2014/A1:2018
Sterilization of health-care products - Ethylene oxide - Requirements for the development validation and routine control of a sterilization process for medical devices - Amendment 1: Revision of Annex E Single batch release
14-529 · Complete
07/15/2019
Not recorded in this export rowSummary [3] · Metadata [15]
Read scope and recognition limits ↗

Prepare for your provider: the proposed protocol/report edition, its justification for your device and the complete current FDA recognition sheet. In particular, inspect exclusions for partial recognition. An edition difference does not automatically make the historical evidence unusable.

Identifier-matching method and selection limits

Match explicit ISO/IEC identifiers in selected preserved summary excerpts to the acquired, count-reconciled FDA metadata export. Preserve the historical edition mention, current snapshot designations, entry dates, recognition extent and any recorded transition-expiration date as separate facts. No title-similarity, edition-equivalence, test-mandate or transition-eligibility inference. At most twenty matched occurrences and thirty metadata rows per reference; wide matches are held for separate review.

8 of 8 matched reference occurrences shown.

Known facts and deciding inputs

Known context: US. Confirm the organisation’s role, relevant activities and markets against the cited criteria.

Source factRecorded valueWhat to check
510(k) identifierK251585 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Generic device categoryProfoject™ Enteral Feeding Syringe [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded applicantCMT Health PTE., Ltd. [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
FDA product codePNR [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Decision date2025-07-18 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded decision codeSESE [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Submission typeTraditional [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Receipt date2025-05-23 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.

Prepare a differences and evidence worksheet

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Indications and users — what differs in our product?

    Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

    Useful evidence: Our proposed indication, user/population and use environment beside this exact recorded indication.

  2. Design and operating principle — what differs in our product?

    Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

    Useful evidence: Design description, materials/components, accessory list and a differences matrix.

  3. Performance evidence actually described — what differs in our product?

    The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

    Useful evidence: The cited narrative and full tables, underlying protocols/results and a reviewed comparison with the proposed product.

  4. Which edition and recognition limits would our provider rely on?

    Compare each historical reference below with the dated FDA metadata. Ask the specialist to justify the applicable scope and selected edition, read the complete supplementary sheet for partial-recognition exclusions, and investigate any recorded transition conditions. A difference in editions does not automatically make the earlier study unusable.

    Useful evidence: Proposed protocol/report edition, dated FDA recognition sheet with scope/exclusions, and a documented edition/transition rationale for the actual device.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — 510(k) decision records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002

  • [1] device-510k-0001-of-0001.json:results[154293] · record 08f561213c7a0da29134c93172bafcc419adc563798a3f163dfada1cfa42897a
  • [16] Complete results array / fda-decision-context:K251585 · record whole dataset partition

FDA 510(k) summary — K251585 ↗

Mon, 04 Aug 2025 16:08:17 GMT · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot ae3e536eb1f3d2e3d64bd06452e455af97e44d13b0bbd64c20ad4df4219445e9

  • [2] PDF page 5 · record 3293eda90a11893cde36c1f0ef0ae2332ec65c5ff95000d055a7b2246be16a79
  • [3] PDF page 6 · record 2a78c151e19542a94a3a7a464b293ad15d62cb1486ff6d33546b4b9c4a6f4627

FDA — recognition metadata exported through the official search form ↗

Version not stated by the source · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 94ccf56dc846c7f6d78d6ad77dc6e87d1803ac3612af01d6894949e34c4a78d8

  • [4] CSV row 820 · record b8108547f36a12d74089b5c6de5000b64de7c362eb1e5e14db7ce0adceeca7d0
  • [5] CSV row 1325 · record 7fa1800eab8c2026e46d10c56f9ce2bbdc5ccb39c35d385a64201c8c280dee1c
  • [6] CSV row 1173 · record 0b157566ec64c539de4b15d5763988d809ea3f3a81078483baad3ed5262b25d2
  • [7] CSV row 118 · record bf8dcf094463a5fcc71bc57cf6e14e9403116f380be6552f39a72da0cbf3e1e6
  • [8] CSV row 1090 · record c19035286c97c432740e6994dce76be1a844df9b8e37f31d18cfd33c09f34131
  • [9] CSV row 1566 · record b830b167e2c493feb8f7b7b83e0f1f4c92b4203eb0c6483590400e66ba23559e
  • [10] CSV row 1567 · record c73693dca274a71a4f2b3140a1825be9f64c28c89899a853e813c1e1d603b617
  • [11] CSV row 212 · record c58ee31140d12898c0db23cda7d00562290acd2368300e246e12d22e784ed97c
  • [12] CSV row 678 · record 6ec28fcf71eda0e35a71b1d0d03543d1f58912431ebda22fcde4f9498888b537
  • [13] CSV row 934 · record 5e5407d514c9a265314228fc33fe24d7475942119240ad6c5284e7de70ccb112
  • [14] CSV row 1133 · record 0e9eb627635be2e73657c9a113fcb24d3df097ff10c5744203a006b1ac13b919
  • [15] CSV row 1134 · record 2ba40dbf47f549b20dd66e6fdd78000e54529b73058ce34fef432c8acca44552

Prepare an editable project brief

Confirm the facts, scope and contact preference before sharing your project. Preparing this page sends no provider outreach.

Choose work packages to discuss

Compare FDA 510(k) Submission Services

Sterilization + Microbiology Validation · Packaging + Shelf-life + Distribution Testing