Recorded device evidence

PodSKAN HF Diagnostic X-ray System (K231761): indications, design and recorded evidence

The acquired K231761 file contains product-specific indications and users, design and operating principle, performance evidence actually described, clinical evidence or its stated absence sections. Compare these dated manufacturer statements with your own indication and design before commissioning evidence work. They describe this submission; they do not prescribe your testing, validate performance independently or establish a suitable predicate.

Evidence retrieved 2026-10-07. Source versions and topic-specific limits are listed below.

Computed exact-code research

This decision beside its recorded product-code cohort

The 2021–2025 complete-year cohort contains 42 selected decisions under primary code KPR. 42 have valid reported receipt/decision pairs; 0 pairs are missing or invalid.

Calendar elapsed days: lower quartile 34.0, median 103.0, upper quartile 156.0. These are recorded calendar differences, not active FDA review time.

The table shows the named decision and up to four latest exact-code research records through 2026-09-27. It can include partial-year records outside the complete-year cohort. The quartiles above use the full stated cohort, not this displayed selection.

Research recordRecorded deviceReceipt → decisionCalendar elapsed daysRecorded submission type
K231761 ↗ · this pagePodSKAN HF Diagnostic X-ray System2023-06-16 → 2023-10-27133Traditional
K260227 ↗EXSYS DEXi (EXSYS DEXi-S40S); EXSYS DEXi (EXSYS DEXi-S40P); EXSYS DEXi (EXSYS DEXi-S40T); EXSYS DEXi (EXSYS DEXi-S50H2026-01-26 → 2026-09-08225Traditional
K260488 ↗MULTIX Impact; MULTIX Impact C2026-02-13 → 2026-08-31199Traditional
K260481 ↗GC85A2026-02-13 → 2026-08-27195Traditional
K260482 ↗DRE Duo2026-02-13 → 2026-06-15122Traditional

Use this comparison: select a record to investigate, read its actual indications and design, and document differences. A shared code or elapsed period cannot establish its relevance to your device.

Calendar elapsed days are not active FDA review time, a promised schedule or a success probability. The selection includes only the stated recorded decisions, not all applications. Latest records can postdate the named historical decision and are contextual research, not its predicates. Shared codes, summaries and observed records do not prove equivalence, current market availability, buyer applicability or predicate suitability.

Calculation, selection and source versions

Reuse the complete-endpoint exact-primary-code, documented SE-decision cohort and its stated cutoff/window. Retain its type-7 quartiles only when at least 20 valid date pairs are present. Independently subtract the named official record’s receipt date from its decision date; missing or negative pairs remain unknown. Show up to four latest exact-code research records, excluding the named ID. Parent analysis hash, source snapshots and calculation versions are retained.

Calculation: fda-named-decision-cohort-context-1; cutoff 2026-09-27; 0 parent records excluded; units record counts; calendar elapsed days.

Source snapshots: 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002.

Read beyond the decision row

Recorded evidence and differences to investigate

The summary is manufacturer-submitted context hosted by FDA. Negative/no-study statements are preserved. Evidence themes are not binding legal requirements.

Short opening excerpts identify the source sections. Read the linked original for the full study context, negative statements, tables and limitations. Public quotations are limited to 120 words per acquired summary; the internal evidence packet retains exact section ranges for review.

Indications and users

K231761 · PDF page 5 ↗ Source 2

PodSKAN       HF  Diagnostic     X-Ray     System    is  intended  for  use  by  the  qualified  and     trained
clinicians  or  radiographers  for  generating  X-rays  of  feet.  It  can  be  used  for  adults  and
paediatrics.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

Design and operating principle

K231761 · PDF page 5 ↗ Source 2

PodSKAN  is  an  advanced  high  frequency  type  X-Ray  system  designed  for  superior  image
quality with powerful 2.8KW generator and very low leakage radiation.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

Performance evidence actually described

K231761 · PDF page 11 ↗ Source 3

The risks identified during risk analysis were reduced by applying suitable risk control
measures and it was noted that there were no unacceptable risks after risk control measures.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

Clinical evidence or its stated absence

K231761 · PDF page 13 ↗ Source 4

Subject device has substantial similarities when compared to the predicate devices.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Preserve whether the summary describes clinical evidence or says none was performed. This recorded case does not determine whether your product needs clinical evidence.

Extraction selection and limits

Select observed whole-line section headings or explicit colon-delimited inline labels on readable PDF pages, including the explicitly documented heading-spacing variants in this parser version. Preserve the original heading, character ranges and layout; do not repair excerpt text. Exclude letter boilerplate and uncertain pages. Stop at observed table headings; do not infer table columns or scalar results from layout text. Multi-column or suspicious split-numeric excerpts are excluded until reviewed alternate extraction is available. Original layout records remain retained; any alternate text has its own extractor-version locator and requires original-page review. For each dimension, prefer usable versioned alternate text where available, selecting its longest usable excerpt; otherwise select the longest usable original-layout excerpt. Ties preserve source order. Bounded to 64 PDF pages and 3,500 characters per section. This is a selection, not a complete dossier analysis.

Excluded extraction items: {}.

Historical references beside dated FDA recognition metadata

Read each mention in the linked original summary, including statements that testing was not performed. Recognition metadata does not determine your testing needs, chosen edition or transition eligibility.

Reference in this historical summaryDesignation in the acquired FDA exportRecognition / entry dateRecorded transition-expiration dateSources
IEC 60601-1:2005
Performance data
IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - Note: This standard is recognized with relevant US national differences applied see references #1 and #2 in the Relevant FDA Guidance and/or Supportive Publication section below.
19-49 · Complete
04/03/2023
Not recorded in this export rowSummary [3] · Metadata [5]
Read scope and recognition limits ↗
IEC 60601-2-28:2010
Performance data
IEC 60601-2-28 Edition 3.0 2017-06
Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
12-309 · Complete
12/23/2019
Not recorded in this export rowSummary [3] · Metadata [6]
Read scope and recognition limits ↗
IEC 60601-2-54:2009
Performance data
IEC 60601-2-54 Edition 2.0 2022-09
Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
12-348 · Partial
12/18/2023
Not recorded in this export rowSummary [3] · Metadata [7]
Read scope and recognition limits ↗
IEC 60601-1-3
Performance data
IEC 60601-1-3 Edition 2.2 2021-01 CONSOLIDATED VERSION
Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
12-336 · Complete
06/07/2021
Not recorded in this export rowSummary [3] · Metadata [8]
Read scope and recognition limits ↗
IEC 60601-1-6:2010
Performance data
IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION
Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
5-132 · Complete
12/21/2020
Not recorded in this export rowSummary [3] · Metadata [9]
Read scope and recognition limits ↗

Prepare for your provider: the proposed protocol/report edition, its justification for your device and the complete current FDA recognition sheet. In particular, inspect exclusions for partial recognition. An edition difference does not automatically make the historical evidence unusable.

Identifier-matching method and selection limits

Match explicit ISO/IEC identifiers in selected preserved summary excerpts to the acquired, count-reconciled FDA metadata export. Preserve the historical edition mention, current snapshot designations, entry dates, recognition extent and any recorded transition-expiration date as separate facts. No title-similarity, edition-equivalence, test-mandate or transition-eligibility inference. At most twenty matched occurrences and thirty metadata rows per reference; wide matches are held for separate review.

7 of 7 matched reference occurrences shown.

Known facts and deciding inputs

Known context: US. Confirm the organisation’s role, relevant activities and markets against the cited criteria.

Source factRecorded valueWhat to check
510(k) identifierK231761 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Generic device categoryPodSKAN HF Diagnostic X-ray System [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded applicantSkanray Technologies Limited [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
FDA product codeKPR [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Decision date2023-10-27 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded decision codeSESE [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Submission typeTraditional [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Receipt date2023-06-16 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.

Prepare a differences and evidence worksheet

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Indications and users — what differs in our product?

    Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

    Useful evidence: Our proposed indication, user/population and use environment beside this exact recorded indication.

  2. Design and operating principle — what differs in our product?

    Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

    Useful evidence: Design description, materials/components, accessory list and a differences matrix.

  3. Performance evidence actually described — what differs in our product?

    The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

    Useful evidence: The cited narrative and full tables, underlying protocols/results and a reviewed comparison with the proposed product.

  4. Clinical evidence or its stated absence — what differs in our product?

    Preserve whether the summary describes clinical evidence or says none was performed. This recorded case does not determine whether your product needs clinical evidence.

    Useful evidence: Your proposed claims and supporting evidence, plus a reviewed rationale for clinical-evidence needs.

  5. Which edition and recognition limits would our provider rely on?

    Compare each historical reference below with the dated FDA metadata. Ask the specialist to justify the applicable scope and selected edition, read the complete supplementary sheet for partial-recognition exclusions, and investigate any recorded transition conditions. A difference in editions does not automatically make the earlier study unusable.

    Useful evidence: Proposed protocol/report edition, dated FDA recognition sheet with scope/exclusions, and a documented edition/transition rationale for the actual device.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — 510(k) decision records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002

  • [1] device-510k-0001-of-0001.json:results[19791] · record 9f71cc1486d947b9207e90b5bb39daf696dfbf5ddee4fa3a6bdec5cac785b5da
  • [10] Complete results array / fda-decision-context:K231761 · record whole dataset partition

FDA 510(k) summary — K231761 ↗

Sun, 29 Oct 2023 22:48:04 GMT · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot df62d0013638437fa2705fb94d3965c1c7db9893f7c8777ec8ea7004da8c513c

  • [2] PDF page 5 · record 5cda72f41326154a90a854d6b005bccf2711dba52341b93d60ae94397f5bc71d
  • [3] PDF page 11 · record 0a132afa25a1251f816fad22998df18aa081f1045d8c935c015f2fa728d18f32
  • [4] PDF page 13 · record 90552176ad7b14045ff791e18a4cdddba1dbbe9e12130c938fe16f065f93b940

FDA — recognition metadata exported through the official search form ↗

Version not stated by the source · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 94ccf56dc846c7f6d78d6ad77dc6e87d1803ac3612af01d6894949e34c4a78d8

  • [5] CSV row 578 · record a8ca2ae93456e86990f02a41106f678908bb91d9543453fd2875c48c4cd817d0
  • [6] CSV row 1070 · record af62fd5fde3e0c40c91f2c3ce32668667d6684f17d246329197e2e3a50b5ff6d
  • [7] CSV row 491 · record f8dfab31257b0854ca481b6f6a2f1f56edc7ef417964f7034789ac30d6c2e741
  • [8] CSV row 836 · record c94fcac3350525c16987153703219902814063008ae8150dc94940f83cfd912b