Recorded device evidence

ImPACT Version 4 (K231688): indications, design and recorded evidence

The acquired K231688 file contains product-specific indications and users, design and operating principle, performance evidence actually described, clinical evidence or its stated absence sections. Compare these dated manufacturer statements with your own indication and design before commissioning evidence work. They describe this submission; they do not prescribe your testing, validate performance independently or establish a suitable predicate.

Evidence retrieved 2026-10-07. Source versions and topic-specific limits are listed below.

Computed exact-code research

This decision beside its recorded product-code cohort

The 2021–2025 complete-year cohort contains 3 selected decisions under primary code POM. 3 have valid reported receipt/decision pairs; 0 pairs are missing or invalid.

No quartiles are displayed: this complete-year cohort has fewer than 20 valid pairs.

The table shows the named decision and up to four latest exact-code research records through 2026-09-27. It can include partial-year records outside the complete-year cohort. The quartiles above use the full stated cohort, not this displayed selection.

Research recordRecorded deviceReceipt → decisionCalendar elapsed daysRecorded submission type
K231688 ↗ · this pageImPACT Version 42023-06-09 → 2023-09-1699Traditional
K241737 ↗Sway System Sports Plus2024-06-17 → 2025-02-15243Traditional
K201376 ↗ANAM Test System2020-05-26 → 2021-03-25303Traditional
K202485 ↗ImPACT Version 42020-08-31 → 2020-12-25116Traditional
K181223 ↗ImPACT2018-05-08 → 2018-10-20165Traditional

Use this comparison: select a record to investigate, read its actual indications and design, and document differences. A shared code or elapsed period cannot establish its relevance to your device.

Calendar elapsed days are not active FDA review time, a promised schedule or a success probability. The selection includes only the stated recorded decisions, not all applications. Latest records can postdate the named historical decision and are contextual research, not its predicates. Shared codes, summaries and observed records do not prove equivalence, current market availability, buyer applicability or predicate suitability.

Calculation, selection and source versions

Reuse the complete-endpoint exact-primary-code, documented SE-decision cohort and its stated cutoff/window. Retain its type-7 quartiles only when at least 20 valid date pairs are present. Independently subtract the named official record’s receipt date from its decision date; missing or negative pairs remain unknown. Show up to four latest exact-code research records, excluding the named ID. Parent analysis hash, source snapshots and calculation versions are retained.

Calculation: fda-named-decision-cohort-context-1; cutoff 2026-09-27; 1 parent records excluded; units record counts; calendar elapsed days.

Source snapshots: 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002.

Read beyond the decision row

Recorded evidence and differences to investigate

The summary is manufacturer-submitted context hosted by FDA. Negative/no-study statements are preserved. Evidence themes are not binding legal requirements.

Short opening excerpts identify the source sections. Read the linked original for the full study context, negative statements, tables and limitations. Public quotations are limited to 120 words per acquired summary; the internal evidence packet retains exact section ranges for review.

Indications and users

K231688 · PDF page 4 ↗ Source 2

ImPACT Version 4 is int ended for use as a comput er -based neurocognit ive t est  bat tery t o aid in t he assessment
and management of concussion.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

Design and operating principle

K231688 · PDF page 4 ↗ Source 2

Read this statement in the original PDF; no complete statement fits the short excerpt.

Compare with your product: Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

Performance evidence actually described

K231688 · PDF page 6 ↗ Source 3

S oftware Verification and                                                    Validation.
   ImPACT Version 4 software was developed, validated, and documented in accordance with IEC 62304 and FDA
   Guidance “General Principles of Software Validation.”

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

Clinical evidence or its stated absence

K231688 · PDF page 6 ↗ Source 3

The 510(k) included the results of clinical studies.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Preserve whether the summary describes clinical evidence or says none was performed. This recorded case does not determine whether your product needs clinical evidence.

Extraction selection and limits

Select observed whole-line section headings or explicit colon-delimited inline labels on readable PDF pages, including the explicitly documented heading-spacing variants in this parser version. Preserve the original heading, character ranges and layout; do not repair excerpt text. Exclude letter boilerplate and uncertain pages. Stop at observed table headings; do not infer table columns or scalar results from layout text. Multi-column or suspicious split-numeric excerpts are excluded until reviewed alternate extraction is available. Original layout records remain retained; any alternate text has its own extractor-version locator and requires original-page review. For each dimension, prefer usable versioned alternate text where available, selecting its longest usable excerpt; otherwise select the longest usable original-layout excerpt. Ties preserve source order. Bounded to 64 PDF pages and 3,500 characters per section. This is a selection, not a complete dossier analysis.

Excluded extraction items: {}.

Historical references beside dated FDA recognition metadata

Read each mention in the linked original summary, including statements that testing was not performed. Recognition metadata does not determine your testing needs, chosen edition or transition eligibility.

Reference in this historical summaryDesignation in the acquired FDA exportRecognition / entry dateRecorded transition-expiration dateSources
IEC 62304
Summary of Performance Testing
IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION
Medical device software - Software life cycle processes
13-79 · Complete
01/14/2019
Not recorded in this export rowSummary [3] · Metadata [4]
Read scope and recognition limits ↗
IEC 62304
Summary of Performance Testing
ANSI AAMI IEC 62304:2006/A1:2016
Medical device software - Software life cycle processes [Including Amendment 1 (2016)]
13-79 · Complete
01/14/2019
Not recorded in this export rowSummary [3] · Metadata [5]
Read scope and recognition limits ↗
ISO 14971
Summary of Performance Testing
ISO 14971 Third Edition 2019-12
Medical devices - Application of risk management to medical devices
5-125 · Complete
12/23/2019
Not recorded in this export rowSummary [3] · Metadata [6]
Read scope and recognition limits ↗
ISO 14971
Summary of Performance Testing
ANSI AAMI ISO 14971: 2019
Medical devices - Applications of risk management to medical devices
5-125 · Complete
12/23/2019
Not recorded in this export rowSummary [3] · Metadata [7]
Read scope and recognition limits ↗

Prepare for your provider: the proposed protocol/report edition, its justification for your device and the complete current FDA recognition sheet. In particular, inspect exclusions for partial recognition. An edition difference does not automatically make the historical evidence unusable.

Identifier-matching method and selection limits

Match explicit ISO/IEC identifiers in selected preserved summary excerpts to the acquired, count-reconciled FDA metadata export. Preserve the historical edition mention, current snapshot designations, entry dates, recognition extent and any recorded transition-expiration date as separate facts. No title-similarity, edition-equivalence, test-mandate or transition-eligibility inference. At most twenty matched occurrences and thirty metadata rows per reference; wide matches are held for separate review.

2 of 2 matched reference occurrences shown.

Known facts and deciding inputs

Known context: US. Confirm the organisation’s role, relevant activities and markets against the cited criteria.

Source factRecorded valueWhat to check
510(k) identifierK231688 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Generic device categoryImPACT Version 4 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded applicantImpact Applications, Inc. [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
FDA product codePOM [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Decision date2023-09-16 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded decision codeSESE [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Submission typeTraditional [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Receipt date2023-06-09 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.

Prepare a differences and evidence worksheet

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Indications and users — what differs in our product?

    Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

    Useful evidence: Our proposed indication, user/population and use environment beside this exact recorded indication.

  2. Design and operating principle — what differs in our product?

    Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

    Useful evidence: Design description, materials/components, accessory list and a differences matrix.

  3. Performance evidence actually described — what differs in our product?

    The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

    Useful evidence: The cited narrative and full tables, underlying protocols/results and a reviewed comparison with the proposed product.

  4. Clinical evidence or its stated absence — what differs in our product?

    Preserve whether the summary describes clinical evidence or says none was performed. This recorded case does not determine whether your product needs clinical evidence.

    Useful evidence: Your proposed claims and supporting evidence, plus a reviewed rationale for clinical-evidence needs.

  5. Which edition and recognition limits would our provider rely on?

    Compare each historical reference below with the dated FDA metadata. Ask the specialist to justify the applicable scope and selected edition, read the complete supplementary sheet for partial-recognition exclusions, and investigate any recorded transition conditions. A difference in editions does not automatically make the earlier study unusable.

    Useful evidence: Proposed protocol/report edition, dated FDA recognition sheet with scope/exclusions, and a documented edition/transition rationale for the actual device.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — 510(k) decision records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002

  • [1] device-510k-0001-of-0001.json:results[21186] · record a86d9423b239951f9b96f08e6ae6f781fe2e88d2aba0f01a90c7904b35fa6bc6
  • [8] Complete results array / fda-decision-context:K231688 · record whole dataset partition

FDA 510(k) summary — K231688 ↗

Mon, 18 Sep 2023 00:03:34 GMT · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 84bf3e03981ff8b84e12ac049cf5b18d21933f2e3710dbc879c04fe860f44a52

  • [2] PDF page 4 · record 47608f22c7a3d819368d35966ef37720800e217d26e87ebf13d2e74952760703
  • [3] PDF page 6 · record 19ba7e610903fb980c87b80e45afa973cba29110d2706ba08a367c6c49e44912

FDA — recognition metadata exported through the official search form ↗

Version not stated by the source · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 94ccf56dc846c7f6d78d6ad77dc6e87d1803ac3612af01d6894949e34c4a78d8

  • [4] CSV row 1228 · record cb5d567e0c94ab972dba3963d8b738f40cec5fdee9610e671a60ec63d6bf4e86
  • [5] CSV row 1229 · record 3177eae036ed6aa78c37390da6cbc7c97c2347c1d13920bbe5c7316a282a51c4
  • [6] CSV row 1011 · record 80abd02faa312db476deb65293af52c610c8ebeac4c377520dd2764555642427
  • [7] CSV row 1012 · record 0425f4fc51db72ca07c5e98184058381b2bb3ee51a538f41f108c706b7a077ec

Prepare an editable project brief

Confirm the facts, scope and contact preference before sharing your project. Preparing this page sends no provider outreach.

Choose work packages to discuss

Compare FDA 510(k) Submission Services

Medical Device Cybersecurity