Recorded device evidence

Quantum Perfusion Dual Lumen Cannula 31F, Quantum Perfusion Dual Lumen Cannula 27F, Quantum Perfusion Dual Lumen Cannula 24F, Quantum Perfusion Dual Lumen Cannula 29F, Quantum Perfusion Dual Lumen Cannula 24F-V1, Quantum Perfusion Dual Lumen Cannula 27F-V1, Quantum Perfusion Dual Lumen Cannula 29F-V1, Quantum Perfusion Dual Lumen Cannula 27F-VS, Quantum Perfusion Dual Lumen Cannula 31F-VS (K221353): indications, design and recorded evidence

The acquired K221353 file contains product-specific indications and users, design and operating principle, recorded comparison and differences, non-clinical evidence actually described, clinical evidence or its stated absence sections. Compare these dated manufacturer statements with your own indication and design before commissioning evidence work. They describe this submission; they do not prescribe your testing, validate performance independently or establish a suitable predicate.

Evidence retrieved 2026-10-07. Source versions and topic-specific limits are listed below.

Computed exact-code research

This decision beside its recorded product-code cohort

The 2021–2025 complete-year cohort contains 16 selected decisions under primary code DWF. 16 have valid reported receipt/decision pairs; 0 pairs are missing or invalid.

No quartiles are displayed: this complete-year cohort has fewer than 20 valid pairs.

The table shows the named decision and up to four latest exact-code research records through 2026-09-27. It can include partial-year records outside the complete-year cohort. The quartiles above use the full stated cohort, not this displayed selection.

Research recordRecorded deviceReceipt → decisionCalendar elapsed daysRecorded submission type
K221353 ↗ · this pageQuantum Perfusion Dual Lumen Cannula 31F, Quantum Perfusion Dual Lumen Cannula 27F, Quantum Perfusion Dual Lumen Cannula 24F, Quantum Perfusion Dual Lumen Cannula 29F, Quantum Perfusion Dual Lumen Cannula 24F-V1, Quantum Perfusion Dual Lumen Cannula 27F-V1, Quantum Perfusion Dual Lumen Cannula 29F-V1, Quantum Perfusion Dual Lumen Cannula 27F-VS, Quantum Perfusion Dual Lumen Cannula 31F-VS2022-05-10 → 2022-08-18100Special
K260195 ↗Elongated One-Piece Arterial (EOPA) 3D™ Arterial Cannulae2026-01-22 → 2026-04-2391Special
K250937 ↗Venous Return Cannulae2025-03-28 → 2026-03-11348Traditional
K260043 ↗Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing Set (SR180C)2026-01-07 → 2026-02-2347Traditional
K253203 ↗Retrograde Coronary Sinus Perfusion Cannulae2025-09-26 → 2026-02-19146Traditional

Use this comparison: select a record to investigate, read its actual indications and design, and document differences. A shared code or elapsed period cannot establish its relevance to your device.

Calendar elapsed days are not active FDA review time, a promised schedule or a success probability. The selection includes only the stated recorded decisions, not all applications. Latest records can postdate the named historical decision and are contextual research, not its predicates. Shared codes, summaries and observed records do not prove equivalence, current market availability, buyer applicability or predicate suitability.

Calculation, selection and source versions

Reuse the complete-endpoint exact-primary-code, documented SE-decision cohort and its stated cutoff/window. Retain its type-7 quartiles only when at least 20 valid date pairs are present. Independently subtract the named official record’s receipt date from its decision date; missing or negative pairs remain unknown. Show up to four latest exact-code research records, excluding the named ID. Parent analysis hash, source snapshots and calculation versions are retained.

Calculation: fda-named-decision-cohort-context-1; cutoff 2026-09-27; 0 parent records excluded; units record counts; calendar elapsed days.

Source snapshots: 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002.

Read beyond the decision row

Recorded evidence and differences to investigate

The summary is manufacturer-submitted context hosted by FDA. Negative/no-study statements are preserved. Evidence themes are not binding legal requirements.

Short opening excerpts identify the source sections. Read the linked original for the full study context, negative statements, tables and limitations. Public quotations are limited to 120 words per acquired summary; the internal evidence packet retains exact section ranges for review.

Indications and users

K221353 · PDF page 14 ↗ Source 2

Read this statement in the original PDF; no complete statement fits the short excerpt.

Compare with your product: Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

Design and operating principle

K221353 · PDF page 13 ↗ Source 3

Read this statement in the original PDF; no complete statement fits the short excerpt.

Compare with your product: Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

Recorded comparison and differences

K221353 · PDF page 14 ↗ Source 2

Read this statement in the original PDF; no complete statement fits the short excerpt.

Compare with your product: The summary’s comparison is a manufacturer statement about that submission. Investigate the actual design differences and cited records before considering any comparator or predicate.

Non-clinical evidence actually described

K221353 · PDF page 15 ↗ Source 4

Read this statement in the original PDF; no complete statement fits the short excerpt.

Compare with your product: Treat these as documented evidence themes in this case, including any limitations or negative statements. Ask which protocols, design inputs and acceptance rationale would be justified for your product; these excerpts are not a mandatory test panel.

Clinical evidence or its stated absence

K221353 · PDF page 15 ↗ Source 4

No clinical data have been included in the current 510(k) submission.

Compare with your product: Preserve whether the summary describes clinical evidence or says none was performed. This recorded case does not determine whether your product needs clinical evidence.

Extraction selection and limits

Select observed whole-line section headings or explicit colon-delimited inline labels on readable PDF pages, including the explicitly documented heading-spacing variants in this parser version. Preserve the original heading, character ranges and layout; do not repair excerpt text. Exclude letter boilerplate and uncertain pages. Stop at observed table headings; do not infer table columns or scalar results from layout text. Multi-column or suspicious split-numeric excerpts are excluded until reviewed alternate extraction is available. Original layout records remain retained; any alternate text has its own extractor-version locator and requires original-page review. For each dimension, prefer usable versioned alternate text where available, selecting its longest usable excerpt; otherwise select the longest usable original-layout excerpt. Ties preserve source order. Bounded to 64 PDF pages and 3,500 characters per section. This is a selection, not a complete dossier analysis.

Excluded extraction items: {'uncertain_pdf_page': 9}.

Historical references beside dated FDA recognition metadata

Read each mention in the linked original summary, including statements that testing was not performed. Recognition metadata does not determine your testing needs, chosen edition or transition eligibility.

Reference in this historical summaryDesignation in the acquired FDA exportRecognition / entry dateRecorded transition-expiration dateSources
ISO 15223-1:2016
NON-CLINICAL TESTING
ISO 15223-1 Fourth edition 2021-07 [Including AMD1: 2025]
Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements [Including Amendment 1 (2025)]
5-148 · Complete
12/22/2025
Not recorded in this export rowSummary [4] · Metadata [5]
Read scope and recognition limits ↗
ISO 15223-1:2016
NON-CLINICAL TESTING
ISO 15223-1 Fourth edition 2021-07
Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
5-134 · Complete
12/20/2021
12/17/2028Summary [4] · Metadata [6]
Read scope and recognition limits ↗
ISO 18193:2021
NON-CLINICAL TESTING
ISO 18193 First edition 2021-08
Cardiovascular implants and artificial organs - Cannulae for extracorporeal circulation
3-176 · Complete
12/20/2021
Not recorded in this export rowSummary [4] · Metadata [7]
Read scope and recognition limits ↗

Prepare for your provider: the proposed protocol/report edition, its justification for your device and the complete current FDA recognition sheet. In particular, inspect exclusions for partial recognition. An edition difference does not automatically make the historical evidence unusable.

Identifier-matching method and selection limits

Match explicit ISO/IEC identifiers in selected preserved summary excerpts to the acquired, count-reconciled FDA metadata export. Preserve the historical edition mention, current snapshot designations, entry dates, recognition extent and any recorded transition-expiration date as separate facts. No title-similarity, edition-equivalence, test-mandate or transition-eligibility inference. At most twenty matched occurrences and thirty metadata rows per reference; wide matches are held for separate review.

2 of 2 matched reference occurrences shown.

Known facts and deciding inputs

Known context: US. Confirm the organisation’s role, relevant activities and markets against the cited criteria.

Source factRecorded valueWhat to check
510(k) identifierK221353 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Generic device categoryQuantum Perfusion Dual Lumen Cannula 31F, Quantum Perfusion Dual Lumen Cannula 27F, Quantum Perfusion Dual Lumen Cannula 24F, Quantum Perfusion Dual Lumen Cannula 29F, Quantum Perfusion Dual Lumen Cannula 24F-V1, Quantum Perfusion Dual Lumen Cannula 27F-V1, Quantum Perfusion Dual Lumen Cannula 29F-V1, Quantum Perfusion Dual Lumen Cannula 27F-VS, Quantum Perfusion Dual Lumen Cannula 31F-VS [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded applicantQura S.R.L [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
FDA product codeDWF [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Decision date2022-08-18 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded decision codeSESE [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Submission typeSpecial [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Receipt date2022-05-10 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.

Prepare a differences and evidence worksheet

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Indications and users — what differs in our product?

    Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

    Useful evidence: Our proposed indication, user/population and use environment beside this exact recorded indication.

  2. Design and operating principle — what differs in our product?

    Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

    Useful evidence: Design description, materials/components, accessory list and a differences matrix.

  3. Recorded comparison and differences — what differs in our product?

    The summary’s comparison is a manufacturer statement about that submission. Investigate the actual design differences and cited records before considering any comparator or predicate.

    Useful evidence: Feature-by-feature differences and the underlying cited documents; no automated predicate selection.

  4. Non-clinical evidence actually described — what differs in our product?

    Treat these as documented evidence themes in this case, including any limitations or negative statements. Ask which protocols, design inputs and acceptance rationale would be justified for your product; these excerpts are not a mandatory test panel.

    Useful evidence: Existing protocols/results and design-specific evidence gaps, with standard editions and scope reviewed separately.

  5. Clinical evidence or its stated absence — what differs in our product?

    Preserve whether the summary describes clinical evidence or says none was performed. This recorded case does not determine whether your product needs clinical evidence.

    Useful evidence: Your proposed claims and supporting evidence, plus a reviewed rationale for clinical-evidence needs.

  6. Which edition and recognition limits would our provider rely on?

    Compare each historical reference below with the dated FDA metadata. Ask the specialist to justify the applicable scope and selected edition, read the complete supplementary sheet for partial-recognition exclusions, and investigate any recorded transition conditions. A difference in editions does not automatically make the earlier study unusable.

    Useful evidence: Proposed protocol/report edition, dated FDA recognition sheet with scope/exclusions, and a documented edition/transition rationale for the actual device.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — 510(k) decision records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002

  • [1] device-510k-0001-of-0001.json:results[150506] · record 8a7a15ff196f186560495e341e2ababa67db7939ce2f149616b5ecd54eb3498d
  • [8] Complete results array / fda-decision-context:K221353 · record whole dataset partition

FDA 510(k) summary — K221353 ↗

Sat, 20 Aug 2022 00:00:06 GMT · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 625e8233c4cf068945043d6bc2ce2d34b9275dc877091a3a7d08bb7c9b59830c

  • [2] PDF page 14 · record 8d260dd8f7fc0fd2bae9155ed906f2315cae67ed160c41a6a9f59dc3fa471a84
  • [3] PDF page 13 · record e7c37e93ab9f3ed977e90816705638c466d007ef2addc6650167e85dec29bd7c
  • [4] PDF page 15 · record 22f2a6a2115072b876fbc408c6c43a6fb0620bb75c853954bdc9b568b9e8c654

FDA — recognition metadata exported through the official search form ↗

Version not stated by the source · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 94ccf56dc846c7f6d78d6ad77dc6e87d1803ac3612af01d6894949e34c4a78d8

  • [5] CSV row 119 · record 4807867f84651019bfe4bf51c9e3ae597bf279825c2b2415648a2657e3903e82
  • [6] CSV row 709 · record 71e42ed9514a35b292e75a119b9fbf800bdf89917ca23000383092ab3ac90bc3
  • [7] CSV row 691 · record 63222abeded0f9d076d31fcd87ccb043af7115e4d13856d78c6163d81543156e

Prepare an editable project brief

Confirm the facts, scope and contact preference before sharing your project. Preparing this page sends no provider outreach.

Choose work packages to discuss

Compare FDA 510(k) Submission Services

Sterilization + Microbiology Validation