Recorded device evidence

Citregen Tendon Interference Screw and Citrelock (K200725): indications, design and recorded evidence

The acquired K200725 file contains product-specific indications and users, design and operating principle, recorded comparison and differences, performance evidence actually described sections. Compare these dated manufacturer statements with your own indication and design before commissioning evidence work. They describe this submission; they do not prescribe your testing, validate performance independently or establish a suitable predicate.

Evidence retrieved 2026-10-08. Source versions and topic-specific limits are listed below.

Computed exact-code research

This decision beside its recorded product-code cohort

The 2021–2025 complete-year cohort contains 37 selected decisions under primary code MAI. 37 have valid reported receipt/decision pairs; 0 pairs are missing or invalid.

Calendar elapsed days: lower quartile 30.0, median 60.0, upper quartile 143.0. These are recorded calendar differences, not active FDA review time.

The table shows the named decision and up to four latest exact-code research records through 2026-09-27. It can include partial-year records outside the complete-year cohort. The quartiles above use the full stated cohort, not this displayed selection.

Research recordRecorded deviceReceipt → decisionCalendar elapsed daysRecorded submission type
K200725 ↗ · this pageCitregen Tendon Interference Screw and Citrelock2020-03-19 → 2020-10-07202Traditional
K262357 ↗Arthrex Bioabsorbable PushLock Suture Anchors2026-07-10 → 2026-08-0728Special
K253931 ↗Arthrex Bioabsorbable PushLock Suture Anchors2025-12-09 → 2026-05-28170Traditional
K260294 ↗SF Push-in Anchor2026-01-29 → 2026-03-2757Traditional
K254055 ↗OSSIOfiber® Suture Anchor2025-12-17 → 2026-02-2469Traditional

Use this comparison: select a record to investigate, read its actual indications and design, and document differences. A shared code or elapsed period cannot establish its relevance to your device.

Calendar elapsed days are not active FDA review time, a promised schedule or a success probability. The selection includes only the stated recorded decisions, not all applications. Latest records can postdate the named historical decision and are contextual research, not its predicates. Shared codes, summaries and observed records do not prove equivalence, current market availability, buyer applicability or predicate suitability.

Calculation, selection and source versions

Reuse the complete-endpoint exact-primary-code, documented SE-decision cohort and its stated cutoff/window. Retain its type-7 quartiles only when at least 20 valid date pairs are present. Independently subtract the named official record’s receipt date from its decision date; missing or negative pairs remain unknown. Show up to four latest exact-code research records, excluding the named ID. Parent analysis hash, source snapshots and calculation versions are retained.

Calculation: fda-named-decision-cohort-context-1; cutoff 2026-09-27; 4 parent records excluded; units record counts; calendar elapsed days.

Source snapshots: 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002.

Read beyond the decision row

Recorded evidence and differences to investigate

The summary is manufacturer-submitted context hosted by FDA. Negative/no-study statements are preserved. Evidence themes are not binding legal requirements.

Short opening excerpts identify the source sections. Read the linked original for the full study context, negative statements, tables and limitations. Public quotations are limited to 120 words per acquired summary; the internal evidence packet retains exact section ranges for review.

Indications and users

K200725 · PDF page 5 ↗ Source 2

Read this statement in the original PDF; no complete statement fits the short excerpt.

Compare with your product: Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

Design and operating principle

K200725 · PDF page 5 ↗ Source 2

The Citregen TIS and Citrelock are fixation devices ranging in diameter from 4 to 9 mm
and  lengths  from  10  to  23  mm  to  be  used  with  reusable  instruments.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

Recorded comparison and differences

K200725 · PDF page 6 ↗ Source 3

The Citregen TIS and Citrelock have the same intended and indications for use, as well
as  similar  technological  characteristics  and  principles  of  operation  as  the  pred               icate
device.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: The summary’s comparison is a manufacturer statement about that submission. Investigate the actual design differences and cited records before considering any comparator or predicate.

Performance evidence actually described

K200725 · PDF page 6 ↗ Source 3

Read this statement in the original PDF; no complete statement fits the short excerpt.

Compare with your product: The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

Extraction selection and limits

Select observed whole-line section headings or explicit colon-delimited inline labels on readable PDF pages, including the explicitly documented heading-spacing variants in this parser version. Preserve the original heading, character ranges and layout; do not repair excerpt text. Exclude letter boilerplate and uncertain pages. Stop at observed table headings; do not infer table columns or scalar results from layout text. Multi-column or suspicious split-numeric excerpts are excluded until reviewed alternate extraction is available. Original layout records remain retained; any alternate text has its own extractor-version locator and requires original-page review. For each dimension, prefer usable versioned alternate text where available, selecting its longest usable excerpt; otherwise select the longest usable original-layout excerpt. Ties preserve source order. Bounded to 64 PDF pages and 3,500 characters per section. This is a selection, not a complete dossier analysis.

Excluded extraction items: {}.

Historical references beside dated FDA recognition metadata

Read each mention in the linked original summary, including statements that testing was not performed. Recognition metadata does not determine your testing needs, chosen edition or transition eligibility.

Reference in this historical summaryDesignation in the acquired FDA exportRecognition / entry dateRecorded transition-expiration dateSources
ISO 10993-18
G. Performance Data
ISO 10993-18 Second edition 2020-01 Amendment 1 2022-05
Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process [Including Amendment 1 (2022)].
2-298 · Partial
09/07/2026
Not recorded in this export rowSummary [3] · Metadata [4]
Read scope and recognition limits ↗

Prepare for your provider: the proposed protocol/report edition, its justification for your device and the complete current FDA recognition sheet. In particular, inspect exclusions for partial recognition. An edition difference does not automatically make the historical evidence unusable.

Identifier-matching method and selection limits

Match explicit ISO/IEC identifiers in selected preserved summary excerpts to the acquired, count-reconciled FDA metadata export. Preserve the historical edition mention, current snapshot designations, entry dates, recognition extent and any recorded transition-expiration date as separate facts. No title-similarity, edition-equivalence, test-mandate or transition-eligibility inference. At most twenty matched occurrences and thirty metadata rows per reference; wide matches are held for separate review.

1 of 1 matched reference occurrences shown.

Known facts and deciding inputs

Known context: US. Confirm the organisation’s role, relevant activities and markets against the cited criteria.

Source factRecorded valueWhat to check
510(k) identifierK200725 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Generic device categoryCitregen Tendon Interference Screw and Citrelock [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded applicantAcuitive Technologies [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
FDA product codeMAI [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Decision date2020-10-07 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded decision codeSESE [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Submission typeTraditional [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Receipt date2020-03-19 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.

Prepare a differences and evidence worksheet

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Indications and users — what differs in our product?

    Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

    Useful evidence: Our proposed indication, user/population and use environment beside this exact recorded indication.

  2. Design and operating principle — what differs in our product?

    Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

    Useful evidence: Design description, materials/components, accessory list and a differences matrix.

  3. Recorded comparison and differences — what differs in our product?

    The summary’s comparison is a manufacturer statement about that submission. Investigate the actual design differences and cited records before considering any comparator or predicate.

    Useful evidence: Feature-by-feature differences and the underlying cited documents; no automated predicate selection.

  4. Performance evidence actually described — what differs in our product?

    The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

    Useful evidence: The cited narrative and full tables, underlying protocols/results and a reviewed comparison with the proposed product.

  5. Which edition and recognition limits would our provider rely on?

    Compare each historical reference below with the dated FDA metadata. Ask the specialist to justify the applicable scope and selected edition, read the complete supplementary sheet for partial-recognition exclusions, and investigate any recorded transition conditions. A difference in editions does not automatically make the earlier study unusable.

    Useful evidence: Proposed protocol/report edition, dated FDA recognition sheet with scope/exclusions, and a documented edition/transition rationale for the actual device.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — 510(k) decision records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002

  • [1] device-510k-0001-of-0001.json:results[103081] · record f9e25772205b1b82c897311126afbdc637326dbda512b25e63d7cffb2327efcb
  • [5] Complete results array / fda-decision-context:K200725 · record whole dataset partition

FDA 510(k) summary — K200725 ↗

Thu, 08 Oct 2020 16:00:10 GMT · retrieved 2026-10-08

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 90ab879f60615c651db230b038678791a0069a4c44bdb6c1ae630f2a09dd0a9c

  • [2] PDF page 5 · record 81e72f773d6dfa05dd4e42514515887d4bfcfc5436b0bb96209ac227edae1aed
  • [3] PDF page 6 · record 567b000e91f2003f2284b361a848f31992ef4f6484230e1afcac234b677365df

Prepare an editable project brief

Confirm the facts, scope and contact preference before sharing your project. Preparing this page sends no provider outreach.

Choose work packages to discuss

Compare FDA 510(k) Submission Services

Biocompatibility Testing (ISO 10993 Program) · Packaging + Shelf-life + Distribution Testing