Recorded device evidence

BEXCORE Breast Biopsy System, Biopsy Needle (K171890): indications, design and recorded evidence

The acquired K171890 file contains product-specific indications and users, design and operating principle, performance evidence actually described sections. Compare these dated manufacturer statements with your own indication and design before commissioning evidence work. They describe this submission; they do not prescribe your testing, validate performance independently or establish a suitable predicate.

Evidence retrieved 2026-10-08. Source versions and topic-specific limits are listed below.

Computed exact-code research

This decision beside its recorded product-code cohort

The 2021–2025 complete-year cohort contains 41 selected decisions under primary code KNW. 41 have valid reported receipt/decision pairs; 0 pairs are missing or invalid.

Calendar elapsed days: lower quartile 106.0, median 158.0, upper quartile 215.0. These are recorded calendar differences, not active FDA review time.

The table shows the named decision and up to four latest exact-code research records through 2026-09-27. It can include partial-year records outside the complete-year cohort. The quartiles above use the full stated cohort, not this displayed selection.

Research recordRecorded deviceReceipt → decisionCalendar elapsed daysRecorded submission type
K171890 ↗ · this pageBEXCORE Breast Biopsy System, Biopsy Needle2017-06-26 → 2018-08-20420Traditional
K260517 ↗Automatic Core Biopsy Instrument2026-02-17 → 2026-08-28192Traditional
K262571 ↗EnCor EnCompass™ Breast Biopsy and Tissue Removal System2026-07-24 → 2026-08-1421Special
K262792 ↗Brevera Biopsy System with CorLumina Imaging Technology (BREVDISP09)2026-08-06 → 2026-08-126Special
K260572 ↗Marrow Marxman2026-02-19 → 2026-08-04166Traditional

Use this comparison: select a record to investigate, read its actual indications and design, and document differences. A shared code or elapsed period cannot establish its relevance to your device.

Calendar elapsed days are not active FDA review time, a promised schedule or a success probability. The selection includes only the stated recorded decisions, not all applications. Latest records can postdate the named historical decision and are contextual research, not its predicates. Shared codes, summaries and observed records do not prove equivalence, current market availability, buyer applicability or predicate suitability.

Calculation, selection and source versions

Reuse the complete-endpoint exact-primary-code, documented SE-decision cohort and its stated cutoff/window. Retain its type-7 quartiles only when at least 20 valid date pairs are present. Independently subtract the named official record’s receipt date from its decision date; missing or negative pairs remain unknown. Show up to four latest exact-code research records, excluding the named ID. Parent analysis hash, source snapshots and calculation versions are retained.

Calculation: fda-named-decision-cohort-context-1; cutoff 2026-09-27; 0 parent records excluded; units record counts; calendar elapsed days.

Source snapshots: 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002.

Read beyond the decision row

Recorded evidence and differences to investigate

The summary is manufacturer-submitted context hosted by FDA. Negative/no-study statements are preserved. Evidence themes are not binding legal requirements.

Short opening excerpts identify the source sections. Read the linked original for the full study context, negative statements, tables and limitations. Public quotations are limited to 120 words per acquired summary; the internal evidence packet retains exact section ranges for review.

Indications and users

K171890 · PDF page 5 ↗ Source 2

The BEXCORE Vacuum-Assisted Breast Biopsy System is indicated to provide breast tissue
 samples for diagnostic sampling of breast abnormalities.

 - It is intended to provide breast tissue for histologic examination with partial or complete
 removal of the imaged abnormality.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

Design and operating principle

K171890 · PDF page 5 ↗ Source 2

BEXCORE Vacuum-Assisted Breast Biopsy (hereinafter referred to as VABB) system is ‘a set of
 equipment used for biopsy’ composed of a sterilized disposable needle probe unit (hereafter Probe)
 and an electronic system unit.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

Performance evidence actually described

K171890 · PDF page 8 ↗ Source 3

Performance testing of the Bexcore Vacuum Assisted Biopsy Needle System was conducted in
 accordance with the following international standards:
 "Guidance on Premarket Notification [51 0(k)] for Medical Devices with Sharps Injury Prevention
 Features; Guidance for Industry and FDA.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

Extraction selection and limits

Select observed whole-line section headings or explicit colon-delimited inline labels on readable PDF pages, including the explicitly documented heading-spacing variants in this parser version. Preserve the original heading, character ranges and layout; do not repair excerpt text. Exclude letter boilerplate and uncertain pages. Stop at observed table headings; do not infer table columns or scalar results from layout text. Multi-column or suspicious split-numeric excerpts are excluded until reviewed alternate extraction is available. Original layout records remain retained; any alternate text has its own extractor-version locator and requires original-page review. For each dimension, prefer usable versioned alternate text where available, selecting its longest usable excerpt; otherwise select the longest usable original-layout excerpt. Ties preserve source order. Bounded to 64 PDF pages and 3,500 characters per section. This is a selection, not a complete dossier analysis.

Excluded extraction items: {'uncertain_pdf_page': 1}.

Historical references beside dated FDA recognition metadata

Read each mention in the linked original summary, including statements that testing was not performed. Recognition metadata does not determine your testing needs, chosen edition or transition eligibility.

Reference in this historical summaryDesignation in the acquired FDA exportRecognition / entry dateRecorded transition-expiration dateSources
ISO 11135:2007
IX. Performance Testing
ISO 11135 Second edition 2014-07-15
Sterilization of health-care products - Ethylene oxide - Requirements for the development validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]
14-529 · Complete
07/15/2019
Not recorded in this export rowSummary [3] · Metadata [4]
Read scope and recognition limits ↗
ISO 11135:2007
IX. Performance Testing
ANSI AAMI ISO 11135:2014/A1:2018
Sterilization of health-care products - Ethylene oxide - Requirements for the development validation and routine control of a sterilization process for medical devices - Amendment 1: Revision of Annex E Single batch release
14-529 · Complete
07/15/2019
Not recorded in this export rowSummary [3] · Metadata [5]
Read scope and recognition limits ↗
ISO 14971:2007
IX. Performance Testing
ISO 14971 Third Edition 2019-12
Medical devices - Application of risk management to medical devices
5-125 · Complete
12/23/2019
Not recorded in this export rowSummary [3] · Metadata [6]
Read scope and recognition limits ↗
ISO 14971:2007
IX. Performance Testing
ANSI AAMI ISO 14971: 2019
Medical devices - Applications of risk management to medical devices
5-125 · Complete
12/23/2019
Not recorded in this export rowSummary [3] · Metadata [7]
Read scope and recognition limits ↗
ISO 10993-11
IX. Performance Testing
ISO 10993-11 Third edition 2017-09
Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
2-255 · Complete
09/17/2018
Not recorded in this export rowSummary [3] · Metadata [8]
Read scope and recognition limits ↗
ISO 10993-11
IX. Performance Testing
ANSI AAMI ISO 10993-11: 2017
Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
2-255 · Complete
09/17/2018
Not recorded in this export rowSummary [3] · Metadata [9]
Read scope and recognition limits ↗
IEC 60601-1-2
IX. Performance Testing
IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
19-36 · Partial
12/21/2020
Not recorded in this export rowSummary [3] · Metadata [10]
Read scope and recognition limits ↗
IEC 60601-1-2
IX. Performance Testing
ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021]
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)]
19-36 · Partial
12/21/2020
Not recorded in this export rowSummary [3] · Metadata [11]
Read scope and recognition limits ↗
IEC 60601-1
IX. Performance Testing
IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - Note: This standard is recognized with relevant US national differences applied see references #1 and #2 in the Relevant FDA Guidance and/or Supportive Publication section below.
19-49 · Complete
04/03/2023
Not recorded in this export rowSummary [3] · Metadata [12]
Read scope and recognition limits ↗

Prepare for your provider: the proposed protocol/report edition, its justification for your device and the complete current FDA recognition sheet. In particular, inspect exclusions for partial recognition. An edition difference does not automatically make the historical evidence unusable.

Identifier-matching method and selection limits

Match explicit ISO/IEC identifiers in selected preserved summary excerpts to the acquired, count-reconciled FDA metadata export. Preserve the historical edition mention, current snapshot designations, entry dates, recognition extent and any recorded transition-expiration date as separate facts. No title-similarity, edition-equivalence, test-mandate or transition-eligibility inference. At most twenty matched occurrences and thirty metadata rows per reference; wide matches are held for separate review.

5 of 5 matched reference occurrences shown.

Known facts and deciding inputs

Known context: US. Confirm the organisation’s role, relevant activities and markets against the cited criteria.

Source factRecorded valueWhat to check
510(k) identifierK171890 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Generic device categoryBEXCORE Breast Biopsy System, Biopsy Needle [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded applicantMedical Park Co., Ltd. [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
FDA product codeKNW [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Decision date2018-08-20 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded decision codeSESE [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Submission typeTraditional [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Receipt date2017-06-26 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.

Prepare a differences and evidence worksheet

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Indications and users — what differs in our product?

    Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

    Useful evidence: Our proposed indication, user/population and use environment beside this exact recorded indication.

  2. Design and operating principle — what differs in our product?

    Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

    Useful evidence: Design description, materials/components, accessory list and a differences matrix.

  3. Performance evidence actually described — what differs in our product?

    The source labels this section as performance data. Determine which described evidence is bench, software, biological or clinical from the actual narrative and original tables; the section label alone does not settle that distinction or prescribe your testing.

    Useful evidence: The cited narrative and full tables, underlying protocols/results and a reviewed comparison with the proposed product.

  4. Which edition and recognition limits would our provider rely on?

    Compare each historical reference below with the dated FDA metadata. Ask the specialist to justify the applicable scope and selected edition, read the complete supplementary sheet for partial-recognition exclusions, and investigate any recorded transition conditions. A difference in editions does not automatically make the earlier study unusable.

    Useful evidence: Proposed protocol/report edition, dated FDA recognition sheet with scope/exclusions, and a documented edition/transition rationale for the actual device.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — 510(k) decision records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002

  • [1] device-510k-0001-of-0001.json:results[144341] · record 111605c3fbcf165d8e0a886b1943102acaf631c35b992538e6af264d6f72cb64
  • [13] Complete results array / fda-decision-context:K171890 · record whole dataset partition

FDA 510(k) summary — K171890 ↗

Thu, 06 Sep 2018 21:35:33 GMT · retrieved 2026-10-08

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot c434322a83755e87a9a4c6ac1022ca92f55383e8478f132868b004115d8661ef

  • [2] PDF page 5 · record 282c356cc8237db9cc2077f849a739e9aadb490560cc51d8bb05ed93fbcb4b68
  • [3] PDF page 8 · record 3dc63f8767a3d60a8bb6865b9a12874d000fce6ecd8571300b81be25b9a97fa8

FDA — recognition metadata exported through the official search form ↗

Version not stated by the source · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 94ccf56dc846c7f6d78d6ad77dc6e87d1803ac3612af01d6894949e34c4a78d8

  • [4] CSV row 1133 · record 0e9eb627635be2e73657c9a113fcb24d3df097ff10c5744203a006b1ac13b919
  • [5] CSV row 1134 · record 2ba40dbf47f549b20dd66e6fdd78000e54529b73058ce34fef432c8acca44552
  • [6] CSV row 1011 · record 80abd02faa312db476deb65293af52c610c8ebeac4c377520dd2764555642427
  • [7] CSV row 1012 · record 0425f4fc51db72ca07c5e98184058381b2bb3ee51a538f41f108c706b7a077ec
  • [8] CSV row 1236 · record e220bfdd467cfe9f0a72cd728a35b729471da73180e93bc747f9b5e0e4734b6f
  • [9] CSV row 1237 · record e4a09c6ce00b3bf6239850d6ca092b48756c1f9b874877abfb67211b290a25da
  • [10] CSV row 876 · record a8f2c7aa5fa6c6ffe1d66d56377f30d7cbd9cdbc2078a3d67e8fdff024e2f0ca
  • [11] CSV row 877 · record 44ad0bacea9459dd513264be920199f91b017f973622203ca12b585f5bc35330
  • [12] CSV row 578 · record a8ca2ae93456e86990f02a41106f678908bb91d9543453fd2875c48c4cd817d0

Prepare an editable project brief

Confirm the facts, scope and contact preference before sharing your project. Preparing this page sends no provider outreach.

Choose work packages to discuss

Compare FDA 510(k) Submission Services

Sterilization + Microbiology Validation · Electrical Safety + EMC Testing (IEC 60601)