Recorded device evidence

Young INFINITY Cordless Handpiece System (K171377): indications, design and recorded evidence

The acquired K171377 file contains product-specific indications and users, design and operating principle, non-clinical evidence actually described sections. Compare these dated manufacturer statements with your own indication and design before commissioning evidence work. They describe this submission; they do not prescribe your testing, validate performance independently or establish a suitable predicate.

Evidence retrieved 2026-10-07. Source versions and topic-specific limits are listed below.

Computed exact-code research

This decision beside its recorded product-code cohort

The 2021–2025 complete-year cohort contains 21 selected decisions under primary code EKX. 21 have valid reported receipt/decision pairs; 0 pairs are missing or invalid.

Calendar elapsed days: lower quartile 179.0, median 246.0, upper quartile 337.0. These are recorded calendar differences, not active FDA review time.

The table shows the named decision and up to four latest exact-code research records through 2026-09-27. It can include partial-year records outside the complete-year cohort. The quartiles above use the full stated cohort, not this displayed selection.

Research recordRecorded deviceReceipt → decisionCalendar elapsed daysRecorded submission type
K171377 ↗ · this pageYoung INFINITY Cordless Handpiece System2017-05-10 → 2018-08-20467Traditional
K261987 ↗Reusable SmartPeg2026-06-12 → 2026-09-0888Traditional
K252223 ↗Alpha Endo Handpiece (Alpha Endo)2025-07-16 → 2025-10-1692Traditional
K251597 ↗InnerView System2025-05-27 → 2025-09-18114Traditional
K243692 ↗EnDrive (EnDriveUS)2024-11-29 → 2025-08-22266Traditional

Use this comparison: select a record to investigate, read its actual indications and design, and document differences. A shared code or elapsed period cannot establish its relevance to your device.

Calendar elapsed days are not active FDA review time, a promised schedule or a success probability. The selection includes only the stated recorded decisions, not all applications. Latest records can postdate the named historical decision and are contextual research, not its predicates. Shared codes, summaries and observed records do not prove equivalence, current market availability, buyer applicability or predicate suitability.

Calculation, selection and source versions

Reuse the complete-endpoint exact-primary-code, documented SE-decision cohort and its stated cutoff/window. Retain its type-7 quartiles only when at least 20 valid date pairs are present. Independently subtract the named official record’s receipt date from its decision date; missing or negative pairs remain unknown. Show up to four latest exact-code research records, excluding the named ID. Parent analysis hash, source snapshots and calculation versions are retained.

Calculation: fda-named-decision-cohort-context-1; cutoff 2026-09-27; 0 parent records excluded; units record counts; calendar elapsed days.

Source snapshots: 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002.

Read beyond the decision row

Recorded evidence and differences to investigate

The summary is manufacturer-submitted context hosted by FDA. Negative/no-study statements are preserved. Evidence themes are not binding legal requirements.

Short opening excerpts identify the source sections. Read the linked original for the full study context, negative statements, tables and limitations. Public quotations are limited to 120 words per acquired summary; the internal evidence packet retains exact section ranges for review.

Indications and users

K171377 · PDF page 6 ↗ Source 2

Battery driven electrical drive unit with wireless foot controller for use with
        disposable prophylaxis angles in hygiene operatory to perform cleaning and
        polishing of tooth surfaces and fillings.











































                                         Confidential and Proprietary
                             Copyright © 2018 Young Dental Manufacturing Co.

Compare with your product: Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

Design and operating principle

K171377 · PDF page 5 ↗ Source 3

The proposed device, the Young INFINITY Cordless Handpiece System, is a
        cordless handpiece which is intended for use by a dental clinician during dental
        cleaning and polishing of tooth surfaces and fillings.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

Non-clinical evidence actually described

K171377 · PDF page 12 ↗ Source 4

Testing included verification of component
        specifications, speed and torque control, connectivity of handpiece / foot
        controller communication, battery life, noise testing, software validation, chemical
        compatibility of housing materials, sterilizability, biocompatibility, fluid ingress,
        and electric al equipment safety.

The excerpt is shortened at a complete statement. Read the complete section in the original PDF before interpreting the evidence.

Compare with your product: Treat these as documented evidence themes in this case, including any limitations or negative statements. Ask which protocols, design inputs and acceptance rationale would be justified for your product; these excerpts are not a mandatory test panel.

Extraction selection and limits

Select observed whole-line section headings or explicit colon-delimited inline labels on readable PDF pages, including the explicitly documented heading-spacing variants in this parser version. Preserve the original heading, character ranges and layout; do not repair excerpt text. Exclude letter boilerplate and uncertain pages. Stop at observed table headings; do not infer table columns or scalar results from layout text. Multi-column or suspicious split-numeric excerpts are excluded until reviewed alternate extraction is available. Original layout records remain retained; any alternate text has its own extractor-version locator and requires original-page review. For each dimension, prefer usable versioned alternate text where available, selecting its longest usable excerpt; otherwise select the longest usable original-layout excerpt. Ties preserve source order. Bounded to 64 PDF pages and 3,500 characters per section. This is a selection, not a complete dossier analysis.

Excluded extraction items: {}.

Historical references beside dated FDA recognition metadata

Read each mention in the linked original summary, including statements that testing was not performed. Recognition metadata does not determine your testing needs, chosen edition or transition eligibility.

Reference in this historical summaryDesignation in the acquired FDA exportRecognition / entry dateRecorded transition-expiration dateSources
ISO 17665-1
Non-Clinical Performance Data
ISO 17665-1 First edition 2006-08-15
Sterilization of health care products - Moist heat - Part 1: Requirements for the development validation and routine control of a sterilization process for medical devices
14-333 · Complete
04/04/2016
07/04/2027Summary [4] · Metadata [5]
Read scope and recognition limits ↗
ISO 17665-1
Non-Clinical Performance Data
ANSI AAMI ISO 17665-1:2006/(R)2013
Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development validation and routine control of a sterilization process for medical devices
14-333 · Complete
04/04/2016
07/04/2027Summary [4] · Metadata [6]
Read scope and recognition limits ↗
IEC 60601-1
Non-Clinical Performance Data
IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - Note: This standard is recognized with relevant US national differences applied see references #1 and #2 in the Relevant FDA Guidance and/or Supportive Publication section below.
19-49 · Complete
04/03/2023
Not recorded in this export rowSummary [4] · Metadata [7]
Read scope and recognition limits ↗
IEC 60601-1-2
Non-Clinical Performance Data
IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
19-36 · Partial
12/21/2020
Not recorded in this export rowSummary [4] · Metadata [8]
Read scope and recognition limits ↗
IEC 60601-1-2
Non-Clinical Performance Data
ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021]
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)]
19-36 · Partial
12/21/2020
Not recorded in this export rowSummary [4] · Metadata [9]
Read scope and recognition limits ↗
ISO 10993-1
Non-Clinical Performance Data
ISO 10993-1 Fifth edition 2018-08
Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
2-258 · Partial
01/14/2019
07/01/2029Summary [4] · Metadata [10]
Read scope and recognition limits ↗
ISO 10993-1
Non-Clinical Performance Data
ANSI AAMI ISO 10993-1: 2018
Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
2-258 · Partial
01/14/2019
07/01/2029Summary [4] · Metadata [11]
Read scope and recognition limits ↗
ISO 10993-1
Non-Clinical Performance Data
ISO 10993-1 Sixth edition 2025-11
Biological evaluation of medical devices - Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process
2-313 · Partial
05/25/2026
Not recorded in this export rowSummary [4] · Metadata [12]
Read scope and recognition limits ↗
ISO 10993-5
Non-Clinical Performance Data
ISO 10993-5 Third edition 2009-06-01
Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
2-245 · Complete
12/23/2016
Not recorded in this export rowSummary [4] · Metadata [13]
Read scope and recognition limits ↗
ISO 10993-5
Non-Clinical Performance Data
ANSI AAMI ISO 10993-5:2009/(R)2014
Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
2-245 · Complete
12/23/2016
Not recorded in this export rowSummary [4] · Metadata [14]
Read scope and recognition limits ↗
ISO 14971
Non-Clinical Performance Data
ISO 14971 Third Edition 2019-12
Medical devices - Application of risk management to medical devices
5-125 · Complete
12/23/2019
Not recorded in this export rowSummary [4] · Metadata [15]
Read scope and recognition limits ↗
ISO 14971
Non-Clinical Performance Data
ANSI AAMI ISO 14971: 2019
Medical devices - Applications of risk management to medical devices
5-125 · Complete
12/23/2019
Not recorded in this export rowSummary [4] · Metadata [16]
Read scope and recognition limits ↗
ISO 14457
Non-Clinical Performance Data
ISO 14457 Second edition 2017-10
Dentistry - Handpieces and motors
4-276 · Partial
06/07/2021
Not recorded in this export rowSummary [4] · Metadata [17]
Read scope and recognition limits ↗

Prepare for your provider: the proposed protocol/report edition, its justification for your device and the complete current FDA recognition sheet. In particular, inspect exclusions for partial recognition. An edition difference does not automatically make the historical evidence unusable.

Identifier-matching method and selection limits

Match explicit ISO/IEC identifiers in selected preserved summary excerpts to the acquired, count-reconciled FDA metadata export. Preserve the historical edition mention, current snapshot designations, entry dates, recognition extent and any recorded transition-expiration date as separate facts. No title-similarity, edition-equivalence, test-mandate or transition-eligibility inference. At most twenty matched occurrences and thirty metadata rows per reference; wide matches are held for separate review.

7 of 7 matched reference occurrences shown.

Known facts and deciding inputs

Known context: US. Confirm the organisation’s role, relevant activities and markets against the cited criteria.

Source factRecorded valueWhat to check
510(k) identifierK171377 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Generic device categoryYoung INFINITY Cordless Handpiece System [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded applicantYoung Dental Manufacturing Co. 1, LLC [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
FDA product codeEKX [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Decision date2018-08-20 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Recorded decision codeSESE [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Submission typeTraditional [1]Dated official record fact; not proof of current commercial availability or applicability to another device.
Receipt date2017-05-10 [1]Dated official record fact; not proof of current commercial availability or applicability to another device.

Prepare a differences and evidence worksheet

Use this checklist to gather your business or product details before speaking with a specialist. The items below explain what to record and suggest useful supporting documents. You can add your own answers in the editable project brief.

  1. Indications and users — what differs in our product?

    Compare the recorded indication, population, body site and use environment with your proposed label. Record each change instead of assuming the primary code settles intended use.

    Useful evidence: Our proposed indication, user/population and use environment beside this exact recorded indication.

  2. Design and operating principle — what differs in our product?

    Compare physical design, materials, energy, accessories and operating principle with your design. Ask which documented differences change the evidence scope.

    Useful evidence: Design description, materials/components, accessory list and a differences matrix.

  3. Non-clinical evidence actually described — what differs in our product?

    Treat these as documented evidence themes in this case, including any limitations or negative statements. Ask which protocols, design inputs and acceptance rationale would be justified for your product; these excerpts are not a mandatory test panel.

    Useful evidence: Existing protocols/results and design-specific evidence gaps, with standard editions and scope reviewed separately.

  4. Which edition and recognition limits would our provider rely on?

    Compare each historical reference below with the dated FDA metadata. Ask the specialist to justify the applicable scope and selected edition, read the complete supplementary sheet for partial-recognition exclusions, and investigate any recorded transition conditions. A difference in editions does not automatically make the earlier study unusable.

    Useful evidence: Proposed protocol/report edition, dated FDA recognition sheet with scope/exclusions, and a documented edition/transition rationale for the actual device.

Work packages and dependencies

What needs to happen first

Questions for providers

Sources and data dates

Read the official document in context. The audit details identify the precise locators and preserved versions used for this page.

FDA openFDA — 510(k) decision records ↗

2026-10-05 · retrieved 2026-10-06

Audit details: precise locators and snapshot identifiers

Source key D01 · snapshot 5a86994ba49a37e71e09be96803fd51981c6c43eaa96caab41aea2b188bfe002

  • [1] device-510k-0001-of-0001.json:results[21856] · record ad1e0b74a89b02542a2123505b5a8dcf7bb88513c3051fed39fe22e9e7b0475c
  • [18] Complete results array / fda-decision-context:K171377 · record whole dataset partition

FDA 510(k) summary — K171377 ↗

Thu, 06 Sep 2018 21:35:30 GMT · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 9d39559336e909540f991fdfc738a9eb4d6be3d750f19eb5316830cfa76b814b

  • [2] PDF page 6 · record 5cd870f6493afce8d53c2b20244d24d901993b1a7c31cc8944855baf66a4f86f
  • [3] PDF page 5 · record 2ef5629463acb18a36bbf99be1b10b6a4dd47ee8a44b9dbb7ea9deb046aa0767
  • [4] PDF page 12 · record 8945dd0b6a836c6c1be4a2897d3bec71fa48b7c9f4f6d7143b62d1eea163e4ae

FDA — recognition metadata exported through the official search form ↗

Version not stated by the source · retrieved 2026-10-07

Audit details: precise locators and snapshot identifiers

Source key D02 · snapshot 94ccf56dc846c7f6d78d6ad77dc6e87d1803ac3612af01d6894949e34c4a78d8

  • [5] CSV row 1553 · record c943a58f778e360c32eefb9f7d93897411ab81b9d9ebcd861ee555120bd72bcc
  • [6] CSV row 1554 · record b09711418b3642a47901d0139325c1f32564bc5a98dab6819be283a8696110f1
  • [7] CSV row 578 · record a8ca2ae93456e86990f02a41106f678908bb91d9543453fd2875c48c4cd817d0
  • [8] CSV row 876 · record a8f2c7aa5fa6c6ffe1d66d56377f30d7cbd9cdbc2078a3d67e8fdff024e2f0ca
  • [9] CSV row 877 · record 44ad0bacea9459dd513264be920199f91b017f973622203ca12b585f5bc35330
  • [10] CSV row 1151 · record eed6cb5e668a5447c86dfed62245d70e5e3e55d6cf3dcf6796391bc4509e3017
  • [11] CSV row 1152 · record 5837b9393097732ea2b59a36e37775bacda9307c0b8de0914383826ebde98568
  • [12] CSV row 4 · record 45cdd48d4d8693839e334ad8c7f9ef9546486f0676625819718db883dc2f6d46
  • [13] CSV row 1428 · record 9b6be60c298d4e35472fc284b7c928b0950ad0e3541bc477011ce5f36cb0e911
  • [14] CSV row 1429 · record 3ea3bbd0d8584755102be0b289ed930495fabbb080ca8cbdc378723cd1d940bb
  • [15] CSV row 1011 · record 80abd02faa312db476deb65293af52c610c8ebeac4c377520dd2764555642427
  • [16] CSV row 1012 · record 0425f4fc51db72ca07c5e98184058381b2bb3ee51a538f41f108c706b7a077ec
  • [17] CSV row 781 · record 7430cf0e398b9b9a4fb068f8c3e59163818393af9c35cbedd9dd0cddae743951